18
dedicated IV set or IV set is installed in the pump already;
Attention: 1. If Air bubble sensor test FAIL, system will show
“
Is the infusion set installed?
”
,
please choose the answer according to the facts, the system will predicate per your choice.
2. If not connect with AC power, but use battery, it will show FAIL, in this situation the pump is
working on battery.
3.Please keep an eye on the self-test, if any abnormal, please contact our company, DO NOT
force to use it.
Press POWER key
for about 2 seconds to turn off the machine.
8.1.3 Fill the IV set and install the IV set properly.
1) Fill the IV set
Connect IV set to infusion bag/bottle and then squeeze the drip chamber to fill with 1/2 of liquid. Open
the flow clip and let the fluid flow to the tip of the needle. Then close the flow clip again.
2
)
Install the IV set
①
Long press the power key
②
Long press OPEN key to open the door