32
Down time
0Hrs 0Min-99Hrs 59Min
8.1.5 Purge
Press Bolus key until the air bubble in infusion tube is totally exhausted, and then close the flow
clamp tightly. The volume generated by “purge” is not calculated in accumulated volume.
Note: “purge” only use
when machine is in non-infusion status, and infusion tube is not
connected to patients.
8.1.6 Start Infusion
Check if parameters are correct and then push “Start” key to start infusion. Only Bolus and Stop key
are available during the infusion except alarm. Under infusion status, the word “Running” will be
displayed at bottom of screen; the indicator light will flash in green.
8.1.7 Bolus(Fast infusion)
There are two methods to purge. They are manual bolus and auto bolus. Please follow by the followin
g procedures:
1)Manual bolus: During infusion process, press bolus key and hold on. The pump will purge at preset
bolus rate. It will return to previous infusion rate after releasing your finger.
2)Auto bolus: After setting bolus rate, it will purge at bolus rate by single clicking bolus key.
Note: Different sizes of infusion sets may have different Bolus rate. Bolus rate can be set in setting int
erface. Please look up 8.4.6.
8.1.8 Stop infusion
Press Stop key to stop the infusion. In this case, the indicator light will be off; the interface will show
volume infused and adjustable parameters.
(1) User could reset the parameters in this interface and press START key to continue infusion.
(2) To complete the infusion, the user could select “End infusion” then press OK key. It will return to
infusion mode setting interface.
8.1.9 Infusion complete
After the preset volume was completed or the accumulated volume reached 36,000ml, the infusion
pump will activate KVO function and give finish alarm. Press Stop key to stop the infusion.