36
Infusion pipe error
alarm
1. Start the device after
installation of hawkmed
dedicated infusion set
Take off the infusion set, then turn off
and turn on the machine to eliminate the
alarm.
★
Note: The infusion pump only has
this function when it works with
hawkmed dedicated infusion set.
E0, E1
E2, E3
1.E0: data communication error
Restart the infusion pump, load the last
infusion setting and start infusion again.
If the alarm is given again, contact with
the manufacturer to inspect and repair.
2. E1: transmission part failure
Restart the infusion pump, load the last
infusion setting and start infusion again.
If the alarm is given again, contact with
the manufacturer to inspect and repair.
3. E2: motor operations faulted
Restart the infusion pump, load the last
infusion setting and start infusion again.
If the alarm is given again, contact with
the manufacturer to inspect and repair.
4. E3: memory fails
Restart the infusion pump. If the alarm is
given again, restore factory defaults and
restart the infusion pump; if the alarm is
still not eliminated, contact with the
manufacturer to inspect and repair.
★
Note: Infusion pipe parameters
should be calibrated again after
restoring factory defaults.