background image

22

Subject to change without notice

V e r t i c a l   s y s t e m

5  Horizontal System (Time Base)

As well as 

time base settings, the horizontal system comprises  

the selection of the trigger position, the zoom functions and the 

available modes of signal capture, the control for the 

marker 

function and the search fcuntions. 

The knobs are used for the adjustment of the time base speed 

and the trigger position. The signal capture modes are selected 

in the respective menus. There is a key provided for activating 

the zoom function.

5.1  Capturing modes RUN and STOP

The capturing modes can be selected with the key RUN/STOP. 

In RUN mode signals will be continuously captured; depending 

on the 

trigger conditions selected, and displayed, erasing the 

previously captured ones. If it is desired to store and further 

analyze a signal and to prevent it being overwritten, capture 

must be stopped by pushing the RUN/STOP key. While in STOP 

mode capture is disabled and the key will light up red.

5.2  Time base adjustments

The large knob in the HORI

-

ZONTAL section of the control 

panel is used for the selection 

of the time base speed. The time 

base speed is displayed in the 

upper left hand corner above 

the graticule. (e.g. „TB:500 ns“) 

To the right there is the display 

of the trigger time position with 

respect to the normal position. 

The normal trigger position is in 

the center of the graticule such 

that 50 % of the signal display 

is before and 50 % is after this 

trigger position. The X POSI

-

TION knob allows continuous 

adjustment of the X position. 

The available maximum values depend on the time base 

setting. By pushing the key SET/CLR the value will be reset 

to its reference position. The arrow keys     

37

 

allow you to 

change the X position by a fixed amount of 5 divisions in the 

respective direction. If marker or search function is chosen 

the arrow keys together with the SET/CLR button are used to 

navigate through and set/clear marker. The key MENU opens 

a menu which allows you to set the X position to its minimum 

and maximum positions or chose the marker function by just 

a key touch. In addition, there is a submenu 

NUMER.INPUT

 

which allows entry of an arbitrary X position. Within this menu 

the search functions can be activated and set. In addition the 

time reference can be set here (position for the trigger point in 

time, from -5 to +5 divisions, 0 is the middle of the screen and 

the standard setting).

5.3  Capture modes

The capture modes are selected by pushing the key ACQUIRE, 

this opens a display menu which offers the basic modes of 

capture:

5.3.1   Refresh: 

In this mode the signals are captured and displayed.

37

D

38

37

41

39

42

43

44

40

Fig. 5.1: Control panel of the 
horizontal system

5.3.2   Roll:

This 

acquisition  mode  is  intended  specifically  for  very  slow 

signals, with the untriggered signal „rolling“ across the screen 

from right to left (requires signals slower than 200 kHz). The 

HMO uses a ring buffer to store the signal values in roll mode. 

Simply put, the instrument writes the first division to the first 

storage space, the second division to the second storage space, 

etc. Once the storage is full, the instrument overwrites the first 

storage space with the data of the most recent measurement 

value. This creates a „ring“ or cycle run, similar to a ticker.

  The ZOOM feature is not available in the roll mode 

(also refer to chapter 5.5 ZOOM Function).

5.3.3:  ENVELOPE:

In this mode, the display includes the normal capture of each 

signal and the maximum as well as the minimum values of 

each capture. Over time, this creates an envelope surrounding 

the signal.

5.3.4: AVERAGE:

In this mode, you can use the 

universal knob in the Cursor/

Menu section of the control panel to set the number of signal 

periods for averaging, available in powers of 2 from 2 to 1024 

(requires repetitive signals).

5.3.5: FILTER:

In this mode, you can activate a low pass filter with adjustable 

cut off frequency to suppress unwanted high frequency inter

-

ferences. The cut off frequency can be set based on the sam

-

pling frequency. The minimum setting is 1/100 of the sampling 

frequency and the maximum value is 1/4 of the sampling rate. 

You can select this setting with the 

universal knob.

The second menu page is accessed by pushing the soft key 

next  to  the  menu  „Page  1I2“,  here,  extended  functions  are 

available:

5.3.6  Peak Detect

This mode is used for very large time base settings to detect 

even short signal changes. You can also deactivate this func-

tion within the menu or you can select the automatic switching 

mode. The following conditions must be met to activate the 

PEAK DETECT

 mode:

–  Function HIGH RESOLUTION is deactivated

–  None of the serial or parallel buses are active

During peak detection, the oscilloscope distinguishes between 

two types: Acquisition peak detection and memory peak de

-

tection.

Each A/D converter converts at the full sampling rate (no 

Interlace mode), even if results have not been written to the 

acquisition memory at full 

sampling rate (for slow time bases, 

for instance). If peak detection is activated, unused converter 

values will be evaluated to detect minimum and maximum 

amplitudes. During this process, the identified minimum and 

maximum values including sampling interval are written to the 

acquisition memory. As a result, the acquisition memory stores 

data pairs representing the signal sequence according to the 

sampling interval. The smallest detectable pulse is the period of 

the maximum sampling rate (no Interlace mode). This describes 

the so-called acquisition peak detection.

A hardware peak detection is not available if data is written 

to the acquisition memory at the ADC‘s maximum sampling 

rate. For slow time bases and a repeat rate set to automatic 

or maximum value, not all data from the acquisition memory 

Summary of Contents for HMO Series

Page 1: ...70 200 MHz Digital Oscilloscope HMO Series 72x 202x Manual English...

Page 2: ...ument have much influence on emission and immunity and therefore on meeting the acceptance limits For different applications the lines and or cables used may be different For measurement operation the...

Page 3: ...Name a channel 21 5 Horizontal System Time Base 22 5 1 Capturing modes RUN and STOP 22 5 2 Time base adjustments 22 5 3 Capture modes 22 5 4 Interlace Mode 24 5 5 ZOOM function 25 5 6 Navigation Funct...

Page 4: ...I UART RS 232 CAN LIN R Automatic Search for User defined Events R Pass Fail Test based on Masks R Vertical Sensitivity 1mV div Offset Control 0 2 20V R 12div x Axis Display Range 20div y Axis Display...

Page 5: ...gic AND OR TRUE FALSE Sources LCH 0 7 CH 1 CH 2 CH 1 CH 4 State LCH 0 7 X H L Duration 8ns 2 147s resolution 8ns Pulses Positive negative Modes equal unequal less than greater than within without a ra...

Page 6: ...Interface IEEE 488 GPIB galvanically isolated HZO91 4RU 19 Rackmount Kit HZO90 Carrying Case for protection and transport HZO20 High voltage probe 1 000 1 400MHz 1 000Vrms HZO30 Active probe 1GHz 0 9p...

Page 7: ...open or in moist atmosphere after any improper transport e g insufficient packing not conforming to the minimum standards of post rail or trans port firm 1 4 Correct operation Please note This instru...

Page 8: ...y class II The measurement category of such probes must be checked and observed The measurement categories were derived corresponding to the distance from the power station and the transients hence to...

Page 9: ...st not be removed from their original packaging until they are ready to be used 4 Keep cells and batteries out of the hands of children If a cell or a battery has been swallowed seek medical aid immed...

Page 10: ...y using the grey soft keys The power key 1 is clearly set apart by its design The most important controls are backlighted by coloured LEDs in order to imme diately indicate the actual settings The pan...

Page 11: ...VGA resolution 640 x 480 pixels In normal mode no menus shown there are 12 divisions in X direction If menus are shown this will be reduced to 10 divi sions On the left of the screen area little arro...

Page 12: ...le and may be installed by the customer in the rear receptacle HO740 IEEE 488 GPIB galvanically isolated HO730 combination of Ethernet and USB with integrated web server All HMO series instruments are...

Page 13: ...g 2 11 Menu for basic settings On the first page you can set the user interface and help language by pressing the soft key LANGUAGE and select Ger man or English The soft key beside MISC opens a menu...

Page 14: ...low frequency compensation and with 1MHz for the high frequency compensation also AUTOMATIC standard setting may be selected In the automatic mode the output will provide 1kHz at sweep speeds from 100...

Page 15: ...oft menu key ACCEPT in order to input the key into the system The option will be activated after restarting the instrument 2 11 Self Alignment The HMO72x 202x series has an internal self alignment in...

Page 16: ...east 3 seconds Fig 3 1 Control panel HMO 3 2 Connection of a probe and signal capture Take one of the probes delivered with the instrument detach the protective cap from the top Apply the compensation...

Page 17: ...how in the top area the complete captured signal below an enlarged portion Use the time base knob to select the zoom factor and the small knob for horizontal positioning Fig 3 6 ZOOM function By press...

Page 18: ...u Press the softmenu key beside MEAS PLACE and choose the number you want using the universal knob You may define the parameter measurement using this menu After switching on the MEASUREMENTS with the...

Page 19: ...ing the respective soft key in this example Screenshots in order to access the settings menu Fig 3 14 Menu Screenshots Please verify that the USB connector into which you plugged the USB stick front o...

Page 20: ...el menu The HMO152x and HMO202x are delivered with the HZO10 a 10 1 probe with automatic attenuation read out which will be read from the probe and factored in The passive probes must be adjusted to t...

Page 21: ...HZ355 probes are recognized by the in strument which automatically selects the appropriate factor If any other probe without automatic recognition of the attenu ation ratio or just a cable is connect...

Page 22: ...ignals with the untriggered signal rolling across the screen from right to left requires signals slower than 200kHz The HMO uses a ring buffer to store the signal values in roll mode Simply put the in...

Page 23: ...illoscope displays a signal window of 600x400 pixels Yt without zoom This translates into 600 data points per detection When peak detection is activated 600 pairs of min max values or 1 200 data value...

Page 24: ...g for display of analogue signals At the linear inter polation there is a straight line used to connect the points Using sample hold type of interpolation allow the exact examination of the position o...

Page 25: ...enerally not possible to zoom in on the memory because the signal values of the X axis are always captured with the maximum memory depth The acquisition mode NORMAL always includes more samples in the...

Page 26: ...period between start and stop slope of the pulse Peak The peak search function searches for pulses within the signal The time of the event is the maximum value of the peak Rise time fall time This fu...

Page 27: ...the SLOPE key occur within the signal selected in the SOURCE menu The signal slope has to pass through the set trigger level The trigger type Slope Trigger is selected in the Autosetup mode AUTOSET k...

Page 28: ...er control panel Press the FILTER key36 then you can select additional settings for the pulse trigger in the soft menu There are 6 different options ti t The pulse width ti is unequal to the reference...

Page 29: ...ith two references t1 and t2 can be set due to pressing the respective soft key and turning the universal knob As with the pulse trigger for ti t or ti t you can set a refer ence time with the soft me...

Page 30: ...cule is divided into horizontal and vertical divisions as CENTER CROSS displays one horizontal and one vertical zero line showing the divisions as dots or as OFF the entire graticule will include no d...

Page 31: ...is will cause the most recent portion of the signal to appear bright while the preceeding portions will fade in proportion to the time elapsed This mode can be selected in the soft menu which will ope...

Page 32: ...termine the settings for the Z input The function SOURCE Z allows you to use any of the analog channels as source for the Z input Use the univer sal knob to select the desired setting The Z input allo...

Page 33: ...IO X This mode provides 3 cursors in order to measure ratios in X direction e g a duty cycle between the first and the second and between the first and the third cursors The values will be presen tedi...

Page 34: ...measurement are available MEAN This mode measures the mean value of the signal amplitude If the signal is periodic the first period on the left of the screen will be used for the measurement The measu...

Page 35: ...e relation of the time delay between the set sources to the signal period of the measurement source This mode searches for the slope of the measurement source that is closest to the time reference The...

Page 36: ...an be used to scale an activated mathematical curve The MATH menu is divided into Quick Mathematics and formula sets Quick Mathematics is designed for simple and quick cal culations The formula sets h...

Page 37: ...The formula editor menu soft menu key MA allows you to activate and deactivate mathematical equations that are de fined and displayed within the selected formula set The list only includes visible eq...

Page 38: ...al knob to choose one of the following operators Addition Subtraction Multiplication Division Maximum Minimum Square Root Amount Pos Wave Neg Wave Reciprocal Inverted Common logarithm Natural logarith...

Page 39: ...ectral lines This function is useful for a precise amplitude measurement of a period signal for instance Square wave The square wave window function multiplies all points by 1 This results in a high f...

Page 40: ...applied to the file when the mask is edited and saved Pressing the soft menu key ACTIONS in the PASS FAIL main menu opens a menu with the available actions The following four actions can be performed...

Page 41: ...a comment which will be displayed in the file manager footer once a file has been selected With the soft key FORMAT and the univer sal knob you can choose the HDS binary data or the SCP plain text for...

Page 42: ...rce for the reference to be saved You can select from the activated channels and mathematical curves Press the soft menu key DISPLAY to display the selected reference curve or to update the current re...

Page 43: ...date Press ACCEPT DIR to confirm the target directory and you will automatically return to the screenshot main menu The soft menu key FILE NAME opens the menu for the name entry where you can use the...

Page 44: ...he HMO generates a 50 Hz or 200 Hz 10 sine wave which feeds the series connection of the test object and a sense resistor If the test object has only a real part such as a resistor both voltages will...

Page 45: ...reen center More important in practice is the quick determination of plain shorts and opens which are the most common causes of requiring service It is highly recommended to observe the necessa ry pre...

Page 46: ...is method all channels may be individually positioned and sized You can reset the position and size of the individual logic chan nels on page 2 2 of the POD menu You also have the option to label the...

Page 47: ...the logic channels were activated some parameters may be measured with the cursors For all activated logic channels of a POD these measurements are available TIME RATIO X V MARKER The results will be...

Page 48: ...ys displayed on the screen to ensure decoding works properly The zoom function allows you to view details for any specific message It is necessary to define a BUS before you can determine the set ting...

Page 49: ...n the BUS short menu 13 2 Parallel BUS The HMO series is able to analyze up to 7 bit lines The soft menu key BUS WIDTH and the universal knob allow you to select the number of bit lines You can use th...

Page 50: ...rt White Stop White No acknowledge Red Acknowledge Green The decoding of the address is performed as a 7 bit value The 8th bit for the write read distinction will be decoded in color not in the HEX va...

Page 51: ...l input use the soft menu key VALUE and the universal knob to set the respective byte value The soft menu key SELECT BYTE allows you to edit the different bytes byte 1 to byte 2 to byte 3 etc sequenti...

Page 52: ...ition to assigning the source the soft menu key ACTIVE allows you to select the following settings CS Chip select high or low active low active is the default setting CLK Data will be stored with risi...

Page 53: ...channels see chapter 4 5 The default setting for both is 500 mV Make sure that a complete message of a serial pro tocol is always displayed on the screen to ensure decoding can function properly The Z...

Page 54: ...t menu key SYMBOL OFFSET and the universal knob are used to select the number of irrelevant sym bols that proceed the pattern within the frame that are relevant for the trigger event Any value between...

Page 55: ...e error OVERLOAD Trigger on CAN Overload frames DATA Trigger on data frames select the correct identifier type via universal knob READ DATA Trigger on read frames select the correct identifier type vi...

Page 56: ...choosing the bus type LIN in the BUS menu and pressing the soft menu key CONFIGURATION In the menu that opens you can press the top soft menu key DATA to select the desired channel via universal knob...

Page 57: ...bit High H 1 Low L 0 or X don t care The state X defines any state If you choose the hexadecimal input the soft menu key VALUE and the universal knob allow you to set the value for the respective byte...

Page 58: ...64 Bit versions The USB interface must be chosen in the oscilloscope and does not need any setting At the first connection Windows ask for a driver The driver you can find on the delivered CD or in th...

Page 59: ...ng modes with slope trigger 27 Fig 6 3 The type B Trigger 28 Fig 6 4 Pulse trigger menu 28 Fig 6 5 Logic trigger menu 28 Fig 6 6 Video trigger menu 29 Fig 7 1 Drawing of the virtual screen area and an...

Page 60: ...ogic channel 29 46 47 50 logic level 46 48 50 52 53 54 56 logic probe 10 12 15 16 27 28 46 48 logic trigger 21 24 28 46 49 M marker function 22 25 mask test 39 40 mathematical functions 18 19 36 37 me...

Page 61: ...trigger signal 11 27 35 trigger source 11 24 27 29 35 trigger type 11 27 29 30 54 two window display 17 U universal knob 10 12 13 15 17 18 19 21 22 25 26 28 29 30 31 32 33 34 35 37 38 39 40 41 42 43...

Page 62: ...62 Subject to change without notice A p p e n d i x...

Page 63: ...63 Subject to change without notice A p p e n d i x...

Page 64: ...er Power Supplies Modular System Series 8000 Programmable Instruments Series 8100 w w w h a m e g c o m 43 2030 2010 43 2030 2010 authorized dealer Oscilloscopes Spectrum Analyzer Power Supplies Modul...

Reviews: