Accessories supplied: Line Cord, Operators Manual on CD-ROM, 2 Probes1:1/ 10:1
40MHz Analog-Oscilloscope HM404
-2
Autoset, Save / Recall, Readout / Cursor and RS-232 Interface
2 Channels, DC - 40MHz, 1mV - 20V/div., Component Tester
Triggering: DC to 100 MHz; Automatic Peak to Peak;
≤≤≤≤≤
0,5div.
Time Base: 0.5 s/div. to 10 ns/div.; with Delay & 2
nd
Trigger.
Specifications
Vertical Deflection
Operating modes:
Channel I or CH II separate,
Channel I and II: alternate or chopped
Chopper Frequency:
approx. 0.5MHz
Sum
or
Difference:
from
Channel I and CH. II
Invert:
CH II
XY-Mode:
via CH I(X) and CH I(Y)
Frequency range:
2x DC to 40MHz (-3dB)
Overshoot:
≤
1%.
Risetime:
<8.75ns
Deflection coefficient:
14 calibrated positions
from
1mV/div to 20V/div
in 1-2-5 sequence,
with variable 2.5:1 up to
50V/div
.
Accuracy in calibrated positions
1mV/div to 2mV/div:
±5% (DC - 10MHz (-3dB))
5mV/div to 20V/div:
±3%
Input impedance:
1M
Ω
II 18pF
Input coupling:
DC - AC - GD (ground)
Input voltage:
max. 400V (DC + peak AC)
Triggering
Automatic
(peak to peak):
≥
0.5div.
Range:
≥
20Hz-100MHz
Normal with level control:
DC-100MHz
≤
(0.5div.)
Indicator for trigger action:
LED
Slope:
positive or negative
Sources:
CH I or II, line, external
ALT. Triggering:
CHI / CHII (
≥
0.8div.)
Coupling:
AC
(10Hz-100MHz)
DC
(0-100MHz)
HF
(50kHz - 100MHz)
LF
(0 - 1.5kHz)
2
nd
triggering:
normal with level control
External:
≥
0.3V
pp
(0 – 100MHz)
Active TV Sync. Separator:
field & line, + / –
Horizontal Deflection
Time coefficients:
22 calibrated steps from
0.5s/div. - 50ns/div. (±3%) in 1-2-5 sequence
Variable 2.5:1 up to 1.25s/div.(uncal.)
X-MAG.x10:
up to
10ns/div.
±5%
Delay:
approx. 140ms - 200ns, variable
Hold-off time:
variable to approx. 10:1
Bandwidth X-amplifier:
0-3MHz (-3dB)
Input X-amplifier:
via Channel I, Sensitivity see CH I
X-Y-phase shift:
<3° below 120kHz.
Operation / Control
Manual
(front panel switches)
Autoset
(automatic parameter selection)
Save/Recall:9
user defined parameter settings
RS232:
interface for remote control via a PC
Readout:
Display of parameter settings
Cursor measurement:
∆
V,
∆
t or
∆
1/t (frequ.)
Component Tester
Test voltage:
approx. 7V
rms
(open circuit)
Test current:
max. 7mA
rms
(short circuit)
Test frequency:
approx.50Hz
One test lead is grounded (Safety Earth)
General Information
CRT:
Screen (8x10cm) internal graticule
Acceleration voltage:
approx
2000V
Trace rotation:
adjustable on front panel
Z Input:
(Intens. modulation), max. +5V (TTL)
Calibrator:
0,2V ±1%,
≈
1kHz/1MHz (t
r
<4ns)
Line voltage:
100-240V AC ±10%, 50/60Hz
Power consumption:
approx. 34 Watt at 50Hz
Min./Max. ambient temperature:
0°C...+40°C
Protective system:
Safety class I (IEC1010-1)
Weight:
approx. 5.6kg.
Color:
techno-brown
Cabinet:
W
285,
H
125,
D
380 mm
Subject to change without notice.
06/98
Signals of 50 and 100 MHz, alternate mode,
display of cursors and frequency values.
TV burst signal in delay mode with 2. trigger.
The excellent user interface characteristics of the new
HM404
-2
oscilloscope are
comparable with high tech scopes. Supported by
two microprocessors
any front
panel input is executed in a fraction of a second. A selftest procedure checks all
relevant parameters of the device; the test results will be displayed on screen within
ten seconds after power on. Supported by an
on screen menu
adjustments can be
performed without opening the scope.
It is recommended to use the
Autoset
function if signals of lower complexity shall
be displayed. The scope’s logic circuitry performs all relevant parameter settings
automatically to optimize the presentation of the signal(s). Of course, any parameter
may be modified manually as required. Front panel settings (measurement parameters)
and selected features are alphanumerically displayed on the screen. The cursor
functions enable the user to analyze a signal while watching the numeric readout for
voltage difference, time difference, or frequency values. Another feature is the sto-
rage capability for nine complete parameter settings, which may be stored and recalled
randomly by pushing the according front panel key.
Because of its high performance characteristics of the broad band signal amplifiers
and its excellent trigger bandwidth the scope is capable to
display 100 MHz
signals.
A delayed time base combined with a second trigger circuit makes the
HM404
-2
an
ideal instrument for high-resolution analysis of expanded, asynchronous signals.
Furthermore, the built in component tester and the
1kHz/1MHz calibrator
are standard
equipment for this class of
HAMEG
scopes.
The instrument may be remotely controlled by any personal computer via its built-
in serial interface. A CD-ROM supplied with the scope, contains the instrument
commands and programming examples.