background image

Subject to change without notice

10

Type of signal voltage

required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.
Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10-3) = 100Hz.
TV-Line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10-5) = 6.4div.
Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x103) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(103x0.2x10-3) = 5div.
Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
required period T = 1:(25x106) = 40ns.

If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(

X-MAG. (x10

) active). In this case, the time interval of interest

can be shifted to the screen center using the

 X-POS. 

control.

When investigating pulse or square waveforms, the critical
feature is the risetime of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the risetime is generally
measured between 10% and 90% of the vertical pulse height.
For measurement, adjust the Y deflection coefficient using its
variable function (uncalibrated) together with the Y-POS.
control so that the pulse height is precisely aligned with the
0% and 100% lines of the internal graticule. The 10% and
90% points of the signal will now coincide with the 10% and
90% graticule lines. The risetime is given by the product of the
horizontal distance in div between these two coincident
points and the calibrated time coefficient setting. The fall time
of a pulse can also be measured by using this method.

The following figure shows correct positioning of the
oscilloscope trace for accurate risetime measurement.

With a time coefficient of 10ns/div (X x10 magnification
active), the example shown in the above figure results in a
total measured risetime of

t

tot

 = 1.6div x 10ns/div = 16ns

When very fast risetimes are being measured, the risetimes
of the oscilloscope amplifier and of the attenuator probe has
to be deducted from the measured time value. The risetime
of the signal can be calculated using the following formula.

t

r

 = 

 t

tot2

 - t

osc2

 - t

p2

In this ttot is the total measured risetime, tosc is the risetime

of the oscilloscope amplifier (approx. 8.75ns), and tp the

risetime of the probe (e.g. = 2ns). If ttot is greater than 100ns,

then ttot can be taken as the risetime of the pulse, and

calculation is unnecessary.

Calculation of the example in the figure above results in a
signal risetime:

         tr

 = √

16

2

 - 8.75

2

 - 2

= 13.25ns

The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
measure in any display position and at any signal amplitude.
It is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness
and that the horizontal distance at 10% and 90% of the
amplitude is measured. If the edge shows rounding or
overshooting, the 100% should not be related to the peak
values but to the mean pulse heights. Breaks or peaks
(glitches) next to the edge are also not taken into account.
With very severe transient distortions, the rise and fall time
measurement has little meaning. For amplifiers with
approximately constant group delay (therefore good pulse
transmission performance) the following numerical relationship
between rise time tr (in ns) and bandwidth B (in MHz) applies:

Connection of Test Signal

In most cases briefly depressing the

 AUTO SET 

causes a

useful signal related instrument setting. The following
explanations refer to special applications and/or signals,
demanding a manual instrument setting.

 The description of

the controls is explained in the section ”controls and
readout”

.

Caution:
When connecting unknown signals to the oscilloscope
input, always use automatic triggering and set the
input coupling switch to AC (readout). The attenuator
should initially be set to 20V/div.

Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient (

VOLTS/DIV.

) in calibrated condition,

an attenuator probe must be inserted before the vertical
input. If, after applying the signal, the trace is nearly blanked,
the period of the signal is probably substantially longer than
the set time deflection coefficient (

TIME/DIV.

). It should be

switched to an adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as

HZ32 

or

 HZ34

, or reduced through a x10 or x100 attenuator

probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to
approx. 50kHz). For higher frequencies, the signal source
must be of low impedance, i.e. matched to the characteristic
resistance of the cable (as a rule 50

). Especially when

transmitting square and pulse signals, a resistor equal to the
characteristic impedance of the cable must also be connected
across the cable directly at the Y-input of the oscilloscope.

Summary of Contents for HM404-2.02

Page 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Page 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Page 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Page 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Page 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Page 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Page 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Page 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Page 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Page 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Page 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Page 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Page 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Page 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Page 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Page 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Page 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Page 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Page 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Page 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Page 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Page 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Page 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Page 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Page 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Page 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Page 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Page 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Page 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Page 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Page 33: ...33 Subject to change without notice Front Panel HM404 2...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: