background image

Subject to change without notice

12

can be used by the operator to save instrument settings
and to recall them. This relates to all settings with the
exception of

 FOCUS, TR 

(trace rotation) and the calibrator

frequency pushbutton.

SAVE:

Press the 

SAVE

 pushbutton briefly to start the save

procedure. The readout then indicates the letter

  ”S”

followed by a cipher between

 1 

and

  9

, indicating the

memory location. If the instrument settings stored in this
memory location must not be overwritten, briefly press
the

 SAVE 

or the

 RECALL 

pushbutton to select another

memory location. Each time the

 SAVE 

pushbutton is

briefly pressed the memory location cipher increases until
the location number 9 is reached. The

 RECALL

pushbutton function is similar but decreases the memory
location cipher until 1 is reached. Press and hold SAVE
for approx. 3 seconds to write the instruments settings
in the memory.

RECALL:

To recall a front panel setup, start that procedure by briefly
pressing the 

RECALL

 pushbutton. The readout then

indicates the letter

 ”R” 

and the memory location number.

If required, select a different memory location as
described above. Recall the settings by pressing and
holding the

 RECALL 

pushbutton for approx. 3 seconds.

Attention:
Make sure that the signal to be displayed is similar to
the one that was present when the settings were
stored. If the signal is different (frequency, amplitude)
to the one during storage then a distorted display may
result.

If the

 SAVE 

or the

 RECALL 

pushbutton was depressed

inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.

Switching the instrument off automatically stores the actual
settings in memory location 9, with the effect that different
settings previously stored in this location get lost. To prevent
this, RECALL 9 before switching the instrument off.

Attention!
Both pushbuttons have a second function if the instru-
ment is switched to menu operation. Please note
”MENU”.

The setting controls and LED’s for the Y amplifiers,
modes, triggering and time base are located under-
neath the sector of the front panel described before.

(8) Y-POS. I

 - Control knob.

The 

vertical trace position

 of channel I can be set with

this control knob. In

 ADD 

(addition) mode both (

Y-POS. I

and

 Y-POS. II

) control knobs are active. If the instrument

is set to

 XY

 mode this control knob is 

inactive

 and the

X-POS. 

knob must be used for a horizontal position shift.

DC voltage measurement:

If no signal is applied at the 

INPUT CHI (26)

, the vertical

trace position represents 0 Volt. This is the case if 

INPUT

CHI (26)

 or in addition 

(ADD)

 mode, both 

INPUT CHI (26)

and 

INPUT CHII (30)

, are set to 

GD

 (ground) and automatic

triggering 

(AT (10))

 is present to make the trace visible.

The trace then can be set to vertical position which is
suited for the following 

DC

 voltage measurement.

Controls and Readout

by the signal‘s pulse duty factor.

In

 DUAL 

mode the

 CURSOR 

lines are related to the signal

which is used for internal triggering.

If the signal height is insufficient, the

 CURSOR 

lines do

not change.

(3) RM

The remote control mode can be switched on or off
(”

RM

” LED dark) via the

 RS232 

interface. On condition

that the ”

RM

” LED is lit, all electronically selectable con-

trols on front panel are inactive. This state can be left by
depressing the 

AUTO SET 

pushbutton provided it was

not deactivated via the interface.

(4) INTENS - READOUT

Knob with associated pushbutton and LEDs.

This control knob is for adjusting the A trace and readout
intensity. Turning this knob clockwise increases and
turning it counterclockwise decreases the intensity.

The

 READOUT 

pushbutton below is for selecting the

function in two ways.

If the readout

 (RO) 

is not switched off, briefly pressing

the

 READOUT 

pushbutton switches over the

 INTENS

knob function indicated by a LED in the sequence:

Yt (time base) mode:

A - RO - A

XY mode:

A - RO - A.

Component  Test:

A - RO - A.

Pressing and holding the

 READOUT 

pushbutton switches

the readout on or off. In readout off condition the

 INTENS

knob function can consequently not be set to

 RO

.

Switching the readout off, may be required if interference
is visible on the signal(s). Such interference may also
originate from the chopper generator if the instrument is
operated in chopped

 DUAL 

mode.

With the exception of the letters ”

CT

” all other

 READOUT

information is switched off in

 COMPONENT TEST 

mode.

All

 INTENS 

settings are stored after the instrument is

switched off.

The

 AUTOSET 

function switches the readout on. The

INTENS

 setting for each function is automatically set to

the mean value, if less intensity was previously selected.

(5) TR

The trace rotation control can be adjusted with a small
screwdriver (

please note ”trace rotation TR”

)

(6) FOCUS

This control knob effects both the trace and the readout
sharpness.

(7) SAVE / RECALL

The instrument contains 9 non volatile memories. These

Summary of Contents for HM404-2.02

Page 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Page 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Page 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Page 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Page 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Page 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Page 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Page 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Page 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Page 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Page 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Page 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Page 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Page 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Page 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Page 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Page 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Page 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Page 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Page 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Page 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Page 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Page 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Page 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Page 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Page 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Page 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Page 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Page 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Page 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Page 33: ...33 Subject to change without notice Front Panel HM404 2...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: