background image

27

Subject to change without notice

Triggering and timebase

Trigger indicator ”TR”

The following description applies to the

  ”TR” LED

.     Please

note item (11) under ”Controls and Readout”.

An LED on condition indicates that the trigger signal has a
sufficient amplitude and the trigger level control setting is
correct. This is valid with automatic and with normal
triggering. By observing the trigger LED, sensitive trigger
level adjustment is possible when normal triggering is
used, particularly at very low signal frequencies. The
indication pulses are of only 100ms duration. Thus for fast
signals the LED appears to glow continuously, for low
repetition rate signals, the LED flashes at the repetition
rate or at a display of several signal periods  not only at the
start of the sweep at the left screen edge, but also at each
signal period.

In automatic triggering mode the sweep generator starts
repeatedly without test signal or external trigger voltage. If
the trigger signal frequency decreases the trigger automatic
repetition frequency the sweep generator starts without
awaiting the trigger pulse. This causes an untriggered
display and a flashing trigger LED.

HOLD OFF-time adjustment

For instrument specific information please note

 DEL.POS.

(22) 

in section ”Controls and Readout”.

If it is found that a trigger point cannot be found on extremely
complex signals, even after careful adjustment of the trigger
level control, a stable display may often be obtained using
the holdoff control. This facility varies the holdoff time
between two sweep periods approx. up to the ratio 10:1.
Pulses or other signal waveforms appearing during this off
period cannot trigger the time base.

Particularly with burst signals or aperiodic pulse trains of
the same amplitude, the start of the sweep can be delayed
until the optimum or required time.

A very noisy signal or a signal with a higher interfering
frequency is at times displayed double. It is possible that
trigger level adjustment only controls the mutual phase
shift, but not the double display. The stable single display
of the signal, required for evaluation, is easily obtainable by
expanding the hold off time until one signal is displayed.

A double display is possible with certain pulse signals,
where the pulses alternately show a small difference of the
peak amplitudes. Only a very exact trigger level adjustment
makes a single display possible. The use of the holdoff
control simplifies the right adjustment.

After specific use the holdoff control should be reset into
its calibration detent (fully ccw), otherwise the brightness
of the display is reduced drastically. The function is shown
in the following figures.

Fig. 1 shows a case where the holdoff control is in the minimum

position and various different waveforms are overlapped
on the screen, making the signal observation unsuccessful.

Fig. 2 shows a case where only the desired parts of the signal are

stably displayed.

Delay / After Delay Triggering

The instrument specific information regarding this mode is
part of the section ”Controls and Readout” paragraph

DEL.POS. (22), SEA./DEL. (24) 

and

 DEL.TRIG. (25)

.

As mentioned before, triggering starts the time base sweep
and unblanks the beam. After the maximum X deflection to
the right, the beam is blanked and flies back to the (left) start
position. After the hold off period the sweep is started
automatically by the automatic trigger or the next trigger
signal. In normal triggering mode the automatic trigger is
switched off and will only start on receipt of a trigger signal.

As the trigger point is always at the trace start position, trace
expansion in X direction with the aid of the time base is limited
to the display on the left of the trace. Parts of the signal to be
expanded which are displayed near the trace end (right side
of the screen) are lost when the time base speed is increased
(time coefficient reduced).

The delay function delays the trace start by a variable time
from the trigger point. This allows the sweep to begin on any
portion of a signal. The time base speed can then be increased
to expand the display in X direction. With higher expansion
rates, the intensity reduces and within certain limits this can
be compensated by the

 INTENS 

knob setting.

If the display shows jitter, it is possible to select for (second)
triggering after the elapsed delay time

 (DTR)

. As mentioned

before, it is possible to display video signals using the frame
sync pulses for triggering

 (TV-F)

. After the delay time set by

the operator, the next line sync pulse or the line content may
be used for triggering. So data lines and test lines can be
displayed separately.

Operation of the delay function is relatively simple. Without
delay function set the time coefficient setting

 (TIME/DIV)

until 1 to 3 signal periods are displayed. Display of less the one
period should be avoided as it limits the selection of the signal
section to be expanded, and may cause trigger problems.

The

 X MAG (x10) 

function should be switched off and the

time variable control should be

 CAL 

position. The signal must

be triggered and stable.

The following explanation assumes that the trace starts on
the left vertical graticule line.

Summary of Contents for HM404-2.02

Page 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Page 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Page 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Page 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Page 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Page 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Page 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Page 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Page 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Page 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Page 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Page 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Page 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Page 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Page 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Page 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Page 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Page 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Page 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Page 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Page 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Page 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Page 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Page 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Page 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Page 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Page 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Page 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Page 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Page 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Page 33: ...33 Subject to change without notice Front Panel HM404 2...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: