background image

23

Subject to change without notice

Operating modes of the vertical amplifiers in Yt mode

Phase comparison between two signals of the same
frequency.

Phase comparison with Lissajous figures

The following diagrams show two sine signals of the same
frequency and amplitude with different phase angles.

Calculation of the phase angle or the phase shift between the
X and Y input voltages (after measuring the distances a and b
on the screen) is quite simple with the following formula, and
a pocket calculator with trigonometric functions. Apart from
the reading accuracy, the signal height has no influence on the
result.

The following must be noted here:

Because of the periodic nature of the trigonometric
functions, the calculation should be limited to angles 

90°

However here is the advantage of the method.

Due to phase shift, do not use a too high test frequency.

It cannot be seen as a matter of course from the screen
display if the test voltage leads or lags the reference
voltage. A CR network before the test voltage input of the
oscilloscope can help here. The 1M

 input resistance can

equally serve as R here, so that only a suitable capacitor C
needs to be connected in series. If the aperture width of
the ellipse is increased (compared with C short-circuited),
then the test voltage leads the reference voltage and vice
versa. This applies only in the region up to 90° phase shift.
Therefore C should be sufficiently large and produce only
a relatively small just observable phase shift.

Should both input voltages be missing or fail in the X-
Y mode, a very bright light dot is displayed on the
screen. This dot can burn into the phosphor at a too high
brightness setting (INTENS. knob) which causes either
a lasting loss of brightness, or in the extreme case,
complete destruction of the phosphor at this point.

Phase difference measurement
in DUAL mode (Yt)

Phase differences between two input signals of the same
frequency and shape can be measured very simply on the
screen in Dual mode. The time base should be triggered by
the reference signal (phase position 0).  The other signal can
then have a leading or lagging phase angle. In alternate
triggering condition, phase difference measurement is not
possible.
For greatest accuracy adjust the time base for slightly over
one period and approximately the same height of both signals
on the screen. The Y deflection coefficients, the time base

coefficients and the trigger level setting can be used for this
adjustment, without influence on the result. Both base lines
are set onto the horizontal graticule center line using the Y-
POS. knobs before the measurement. With sinusoidal signals,
observe the zero (crossover point) transitions; the sine peaks
are less accurate. If a sine signal is noticeably distorted by
even harmonics, or if a DC voltage is present, AC coupling is
recommended for both channels. If it is a question of pulses
of the same shape, read off at steep edges.

It must be noted that the phase difference cannot be
determined if alternate triggering is selected.

Phase difference measurement in DUAL mode

t

= horizontal spacing of the zero transitions in div

T = horizontal spacing for one period in div

In the example illustrated, t = 3div and T = 10div The phase
difference in degrees is calculated from

or expressed in radians

Relatively small phase angles at not too high frequencies can
be measured more accurately in the X-Y mode with Lissajous
figures.

Measurement of an amplitude modulation

The momentary amplitude u at time t of a HF-carrier voltage,
which is amplitude modulated without distortion by a sinusoidal
AF voltage, is in accordance with the equation

where

U

T

 = unmodulated carrier amplitude

  = 2

π

F = angular carrier frequency

ω

   = 2

π

f = modulation angular frequency

m   = modulation factor.

The lower side frequency F-f  and the upper side frequency
F+f  arise because of the modulation apart from the carrier
frequency F.

The display of an amplitude-modulated HF oscillation can be
evaluated with the oscilloscope provided the frequency
spectrum is inside the oscilloscope bandwidth. The time base
is set so that several cycles of the modulation frequency are
visible. Strictly speaking, triggering should be external with
modulation frequency (from the AF generator or a demo-
dulator). However, internal triggering is frequently possible
with normal triggering using a suitable trigger level setting and
possibly also using the time vernier (variable) adjustment.

Summary of Contents for HM404-2.02

Page 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Page 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Page 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Page 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Page 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Page 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Page 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Page 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Page 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Page 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Page 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Page 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Page 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Page 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Page 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Page 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Page 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Page 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Page 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Page 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Page 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Page 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Page 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Page 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Page 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Page 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Page 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Page 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Page 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Page 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Page 33: ...33 Subject to change without notice Front Panel HM404 2...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: