background image

Subject to change without notice

30

Component Tester

Component Tester

General

The instrument specific information regarding the control and
terminals are part of item

 (39) 

in section ”Controls and Readout”.

The instrument has a built-in electronic Component Tester,
which is used for instant display of a test pattern to indicate
whether or not components are faulty. It can be used for quick
checks of semiconductors (e.g. diodes and transistors),
resistors, capacitors, and inductors. Certain tests can also be
made to integrated circuits. All these components can be
tested individually, or in circuit provided that it is unpowered.

The test principle is fascinatingly simple. A built-in generator
delivers a sine voltage, which is applied across the component
under test and a built-in fixed resistor. The sine voltage across
the test object is used for the horizontal deflection, and the
voltage drop across the resistor (i.e. current through test
object) is used for vertical deflection of the oscilloscope. The
test pattern shows a current-voltage characteristic of the test
object.

The measurement range of the component tester is limited
and depends on the maximum test voltage and current
(please note data sheet). The impedance of the component
under test is limited to a range from approx. 20

 to 4.7k

.

Below and above these values, the test pattern shows only
short-circuit or open-circuit. For the interpretation of the
displayed test pattern, these limits should always be born in
mind. However, most electronic components can normally be
tested without any restriction.

Using the Component Tester

After the component tester is switched on, the vertical
preamplifier and the time base generator are inoperative. A
shortened horizontal trace will be observed. It is not necessary
to disconnect scope input cables unless in-circuit
measurements are to be carried out.

For the component connection, two simple test leads with
4mm  Ø banana plugs, and with test prod, alligator clip or
sprung hook, are required. The test leads are connected as
described in section ”Controls and Readout”.

Test Procedure

Caution!
Do not test any component in live circuitry -  remove all
grounds, power and signals connected to the compo-
nent under test. Set up Component Tester as stated.
Connect test leads across component to be tested.
Observe oscilloscope display.

Only discharged capacitors should be tested!

Test Pattern Displays

The page ”Test patterns” shows typical patterns displayed by
the various components under test.
• Open circuit is indicated by a straight horizontal line.
• Short circuit is shown by a straight vertical line.

Testing Resistors

If the test object has a linear ohmic resistance, both deflecting
voltages are in the same phase. The test pattern expected

from a resistor is therefore a sloping straight line. The angle
of slope is determined by the resistance of the resistor under
test. With high values of resistance, the slope will tend
towards the horizontal axis, and with low values, the slope will
move towards the vertical axis. Values of resistance from 20

to 4.7k

 can be approximately evaluated. The determination

of actual values will come with experience, or by direct
comparison with a component of a known value.

Testing Capacitors and Inductors

Capacitors and inductors cause a phase difference between
current and voltage, and therefore between the X and Y
deflection, giving an ellipse-shaped display. The position and
opening width of the ellipse will vary according to the
impedance value (at 50Hz) of the component under test.

A horizontal ellipse indicates a high impedance or a relatively
small capacitance or a relatively high inductance.

A vertical ellipse indicates a small impedance or a relatively
large capacitance or a relatively small inductance.

A sloping ellipse means that the component has a considerable
ohmic resistance in addition to its reactance.

The values of capacitance of normal or electrolytic capacitors
from 0.1µF to 1000µF can be displayed and approximate
values obtained. More precise measurement can be obtained
in a smaller range by comparing the capacitor under test with
a capacitor of known value. Inductive components (coils,
transformers) can also be tested. The determination of the
value of inductance needs some experience, because inductors
have usually a higher ohmic series resistance. However, the
impedance value (at 50Hz) of an inductor in the range from
20

 to 4.7k

 can easily be obtained or compared.

Testing Semiconductors

Most semiconductor devices, such as diodes, Z-diodes,
transistors, FETs can be tested. The test pattern displays vary
according to the component type as shown in the figures
below. The main characteristic displayed during semiconductor
testing is the voltage dependent knee caused by the junction
changing from the conducting state to the non conducting
state. It should be noted that both the forward and the reverse
characteristic are displayed simultaneously. This is a two-
terminal test, therefore testing of transistor amplification is
not possible, but testing of a single junction is easily and
quickly possible. Since the test voltage applied is only very
low, all sections of most semi-conductors can be tested
without damage. However, checking the breakdown or reverse
voltage of high voltage semiconductors is not possible. More
important is testing components for open or short-circuit,
which from experience is most frequently needed.

Testing Diodes

Diodes normally show at least their knee in the forward
characteristic. This is not valid for some high voltage diode
types, because they contain a series connection of several
diodes. Possibly only a small portion of the knee is visible.
Zener diodes always show their forward knee and, depending
on the test voltage, their zener breakdown forms a second
knee in the opposite direction. If the breakdown voltage is
higher than the positive or negative voltage peak of the test
voltage, it can not be displayed.

Summary of Contents for HM404-2.02

Page 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Page 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Page 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Page 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Page 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Page 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Page 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Page 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Page 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Page 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Page 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Page 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Page 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Page 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Page 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Page 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Page 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Page 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Page 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Page 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Page 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Page 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Page 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Page 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Page 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Page 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Page 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Page 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Page 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Page 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Page 33: ...33 Subject to change without notice Front Panel HM404 2...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: