Software
67
The chapter 9.1.3 Heater & Cell Status describes how a liquid test cell could be assigned to a
liquid heater and gives more information about the test cells status. The following liquid and solid
heater and cell status field gives a basic description:
Selected Liquid Test Cell
The yellow bar indicates the selected liquid test
cell. This cell will be measured.
▪
Liquid heater #2 is selected and test cell 2 is
assigned to it.
▪
“File_2” is selected for the report and results
▪
Heating is switched on
▪
Set temperature is 90
C and the actual
temperature is 131.0
C
▪
Measurement is running (yellow warning
symbol)
Selected Solid Test Cell
The yellow bar indicates the selection of the solid
test cell. This cell will be measured.
Solid heater & test cell are selected.
“File_3” is selected for the report and results
Heating is switched on
Set temperature is 100
C and the actual
temperature is 0.1
C
Measurement is running (yellow warning symbol)
After starting the DC Resistivity measurement every 2 seconds a measuring value will be displayed
in the following measurement value section. Each of this values will also be stored in the result
table. After the usual set measuring time of 1 minute (according to the standards) the
measurement will automatically stops and the DUT will be discharged. The last value after 1
minute is the result of the DC Resistivity measurement.
In the measurement value section the actual applied DC voltage, the DC Current, the Resistivity,
the Ambient Temperature and the relative Humidity are displayed. The last measured values will
not be cleared until a new DC Resistivity measurement starts or a shorting phase is started.