104
Troubleshooting
15
Troubleshooting
All error messages appear on the display of the measuring instrument. If persistent problems or
faulty operation should occur then please contact the Customer Support Department of HAEFELY
TEST AG or your local agent.
The Customer Support Department can be reached at the following postal address:
HAEFELY TEST AG
Customer Service - Tettex
Birsstrasse 300
CH-4052 Basel
Switzerland
Tel:
+41 61 373 4422
Fax:
+41 61 373 4914
e-mail:
15.1
Windows Recovery
The 2830 has an integrated PC board and runs on Windows Embedded 7. In case of a damaged
Windows operating system (corrupted files, damaged partition etc.) the integrated Windows
recovery function can help to restore the instrument in the state it was delivery from the factory.
There
are
two
drives
available:
C:/
Contains
the
Windows
installation
and
all
system
files
D:/
Contains the user data and installation files of the application software
During the recovery process all files, software, programs, etc. saved on partition C:/ will
be deleted.
Any update of the 2830 software that was installed after delivery have to be reinstalled
after the recovery process