Software
65
9.4.1.3
Log
This menu records all measured data in function of the time. It displays some statistical data as
average and standard deviation. Be aware that the recorded data looks very instable. This is
caused by the automatic scaling of the scope, where the lowest and the highest value will be used
for minimum and maximum y-axis. At maximum 1000 values are recorded.
Buttons description:
X / Y Axis
The measuring values for the x- or y-axis can be chosen.
Clear Curve
Clear recorded values.
Auto Scale
Enable/disable automatic scaling of the scope. If it is unchecked
higher and lower values than the maximum range of the scope are
not shown anymore
Recording
If this option is deactivated, the recording of the measuring values will
be stopped.