
40
Software
By pressing this button,
you can select between “Exit to
Windows” and “Shut Down”.
Click the “Exit to Windows” button to terminate the
application software and exit to the Windows Operating
System.
Click the “Shut Down” button to terminate the software and
shut down the system.
It’s strongly recommended to shut down the system correctly
before switching the main power off.
9.1.6
Alarm Messages
Following alarm messages could be displayed in the title bar (see chapter 9.1.5 Title Bar):
Interlock #1
The interlock of the liquid test cell 1 is open.
Close the cover (if available) of the test cell or connect the interlock plug to the interlock 1 plug
socket on the rear side of the 2831.
Interlock #2
The interlock of the liquid test cell 2 is open.
Close the cover (if available) of the test cell or connect the interlock plug to the interlock 2 plug
socket on the rear side of the 2831.
Emergency
The emergency button on the front panel of the 2831 is pressed.
Unlock the button by turning it clockwise.
Emergency or Interlock not connected
When this alarm messages will be displayed then the emergency button on the front panel is
pressed and one or both interlock plugs are not connected.
Unlock the emergency button by turning it clockwise and check both interlock plugs or close the
cover (if existing) of the test cells.
2830 not connected
Summary of Contents for 2830
Page 1: ...2830 2831 Precision Liquid and Solid Dielectric Analyzer User Manual 4843477 M Version 3 0 ...
Page 7: ...Introduction VII 17 Conformity 113 18 Notes 114 ...
Page 8: ...VIII Introduction ...
Page 121: ...Conformity 113 17 Conformity ...
Page 122: ...114 Notes 18 Notes ...
Page 123: ...Notes 115 ...