Option
Description
SELECT
PARAM.
Customizes the options for sensor data handling and
storage. Refer to
on page 11.
RESET
DEFAULTS
Sets the configuration menu to the default settings. All
sensor information is lost.
Select the parameters
1.
Select the type of chlorine sensor used - Total CL2 or Free CL2.
2.
Select whether a pH sensor is used - Yes or No.
3.
If Yes, select the type of pH sensor used - DIFF PH (pHD) or
COMBO pH (pH combination) and then Chlorine.
4.
Customize the sensor parameters:
Option
Description
SELECT
UNITS
Sets the units for the sensor measurements-Auto ppb-ppm,
Auto ug/L-mg/L, Fixed ppm or Fixed mg/L.
DISPLAY
FORMAT
Sets the number of decimal places that are shown on the
measure screen-X.XXX, XX.XX (default), XXX.X or XXXX
(Auto).
SELECT
RANGE
Sets the measurement range-0 to 10 ppm.
CAL WATCH
Shown if pH sensor is used-refer to
on page 11.
FILTER
Sets a time constant to increase signal stability. The time
constant calculates the average value during a specified
time-0 (filtering disabled) to 60 seconds (average of signal
value for 60 seconds). The filter increases the time for the
sensor signal to respond to changes in the process.
LOG SETUP
Sets the time interval for event and data logging for
chlorine concentration and flow status-10, 30 seconds, 1,5,
15 (default), 60 minutes.
Select the Cal Watch alarm values
The Cal Watch menu is used to:
• Set the error and warning alarm conditions for chlorine and/or pH
measurement deviations.
• Set the amount of time a chlorine and/or pH measurement can be
outside the deviation range before an alarm occurs.
• Set the amount of time a chlorine measurement can be 0.5 ppm or
higher before an alarm occurs if the previous chlorine sensor
calibration was done using a process flow with a low chlorine
concentration (LCC) (< 0.5 ppm).
• Set the amount of time a Cal Watch alarm is on before it is canceled
by the instrument if measurements go back in to range.
• Set the percentage rate at which measurements must be outside the
deviation range before an alarm occurs and must be back in range
before an alarm is canceled by the instrument.
To select Cal Watch alarm values:
1.
Select Cal Monitor.
2.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
3.
Select the measurement(s) to monitor (choose one):
Option
Description
ALL
Enables an error or warning alarm to activate when chlorine
and/or pH measurement deviations occur that are equal to or
higher than the chlorine and pH deviation values selected by
the user.
CL2 ONLY
Enables an error or warning alarm to activate when a chlorine
measurement deviation occurs that is equal to or higher than
the chlorine measurement deviation values selected by the
user.
pH ONLY
Enables an error or warning alarm to activate when a pH
measurement deviation occurs that is equal to or higher than
the pH measurement deviation values selected by the user.
None
Disables all Cal Watch alarms.
English
11
Summary of Contents for Chlorine Sensor
Page 8: ...8 English...
Page 17: ...English 17...
Page 29: ...Deutsch 29...
Page 51: ...Assemblare il sensore per cloro Italiano 51...
Page 53: ...Italiano 53...
Page 73: ...Assemblage du capteur de chlore Fran ais 73...
Page 75: ...Fran ais 75...
Page 95: ...Montaje del sensor de cloro Espa ol 95...
Page 97: ...Espa ol 97...
Page 105: ...Espa ol 105...
Page 117: ...Montar o sensor de cloro Portugu s 117...
Page 119: ...Portugu s 119...
Page 139: ...e tina 139...
Page 160: ...160 Dansk...
Page 181: ...Chloorsensor samenbouwen Nederlands 181...
Page 183: ...Nederlands 183...
Page 203: ...Monta czujnika chloru Polski 203...
Page 205: ...Polski 205...
Page 225: ...Svenska 225...
Page 235: ...Svenska 235...
Page 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Page 244: ...1 1 4 2 5 3 6 30 mV 244...
Page 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Page 246: ...48 246...
Page 247: ...247...
Page 248: ...6 12 260 248...
Page 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Page 257: ...6 1 3 6 3 1 1 6 257...
Page 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Page 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Page 270: ...270 Magyar...
Page 280: ...280 Magyar...
Page 291: ...Asamblarea senzorului de clor Rom n 291...
Page 293: ...Rom n 293...
Page 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Page 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Page 311: ...OFF 30 pH pH 2 311...
Page 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Page 313: ...313...
Page 314: ...6 12 327 314...
Page 315: ...315...
Page 324: ...1 1 324...
Page 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Page 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Page 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Page 337: ...T rk e 337...
Page 358: ...358 Slovensk jazyk...
Page 368: ...368 Slovensk jazyk...
Page 379: ...Slovenski 379...
Page 400: ...400 Hrvatski...
Page 417: ...1 2 PVC 1 2 417...
Page 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Page 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Page 420: ...MSDS M48 M48 O ring 420...
Page 421: ...421...
Page 422: ...6 12 435 422...
Page 423: ...423...
Page 432: ...1 1 432...
Page 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Page 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Page 440: ......
Page 441: ......