6.
Assemble the sensor with a new membrane cap.
7.
Install the sensor in the flow cell and connect the sensor cable.
8.
Condition the sensor by operating it for 6 to 12 hours. The sensor is
conditioned when the sensor readings stabilize. Refer to
on page 20 for information about viewing sensor
readings.
9.
Calibrate the sensor.
Replace the electrolyte
1.
Disconnect the cable from the sensor.
2.
Remove the sensor from the flow cell.
3.
Read the precautions in
on page 7.
4.
Lift up the rubber band that covers the vent hole marked “M48” on
the membrane cap and slide the rubber band over so it does not
cover the vent hole.
5.
Turn the membrane cap counter-clockwise and remove the
membrane cap from the sensor.
6.
Hold the membrane cap with the electrolyte on the bottom and shake
up and down 1-2 times to remove old electrolyte.
7.
Assemble the sensor with the membrane cap.
8.
Install the sensor in the flow cell and connect the sensor cable.
9.
Condition the sensor by operating it until the sensor readings
stabilize. Refer to
on page 20 for
information about viewing sensor readings.
10.
Calibrate the sensor.
Prepare the sensor for storage
C A U T I O N
The electrolyte contains potassium halide and buffer to adjust acidity. Read the
MSDS sheet before opening the electrolyte bottle.
The sensor can be stored for up to 3 years from the date of manufacture
when prepared as described. Refer to
on page 3 for
storage specifications.
Note: Do not touch the electrodes and keep the electrodes clean. Do not remove
the layer on the electrodes.
1.
Turn the membrane cap counter-clockwise and remove the
membrane cap from the sensor.
2.
Rinse the membrane cap and electrode in clean water.
3.
Dry the membrane cap and electrode with a clean, dry paper towel in
a place free of dust.
4.
Loosely put the membrane cap on to the electrode shaft to protect
the electrode. The membrane must not rest against the measuring
electrode.
Remove the sensor from storage
Pre-requisite:
Obtain a new membrane cap. The old membrane cap
cannot be used again.
1.
on page 7.
2.
Turn the membrane cap counter-clockwise and remove the
membrane cap from the sensor.
3.
Dispose of the old membrane cap.
4.
Polish the electrode tip with the special abrasive paper.
5.
Assemble the sensor with a new membrane cap.
6.
Install the sensor in the flow cell and connect the sensor cable.
7.
Condition the sensor by operating it for 6 to 12 hours. The sensor is
conditioned when the sensor readings stabilize. Refer to
on page 20 for information about viewing sensor
readings.
8.
Calibrate the sensor.
English
19
Summary of Contents for Chlorine Sensor
Page 8: ...8 English...
Page 17: ...English 17...
Page 29: ...Deutsch 29...
Page 51: ...Assemblare il sensore per cloro Italiano 51...
Page 53: ...Italiano 53...
Page 73: ...Assemblage du capteur de chlore Fran ais 73...
Page 75: ...Fran ais 75...
Page 95: ...Montaje del sensor de cloro Espa ol 95...
Page 97: ...Espa ol 97...
Page 105: ...Espa ol 105...
Page 117: ...Montar o sensor de cloro Portugu s 117...
Page 119: ...Portugu s 119...
Page 139: ...e tina 139...
Page 160: ...160 Dansk...
Page 181: ...Chloorsensor samenbouwen Nederlands 181...
Page 183: ...Nederlands 183...
Page 203: ...Monta czujnika chloru Polski 203...
Page 205: ...Polski 205...
Page 225: ...Svenska 225...
Page 235: ...Svenska 235...
Page 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Page 244: ...1 1 4 2 5 3 6 30 mV 244...
Page 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Page 246: ...48 246...
Page 247: ...247...
Page 248: ...6 12 260 248...
Page 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Page 257: ...6 1 3 6 3 1 1 6 257...
Page 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Page 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Page 270: ...270 Magyar...
Page 280: ...280 Magyar...
Page 291: ...Asamblarea senzorului de clor Rom n 291...
Page 293: ...Rom n 293...
Page 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Page 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Page 311: ...OFF 30 pH pH 2 311...
Page 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Page 313: ...313...
Page 314: ...6 12 327 314...
Page 315: ...315...
Page 324: ...1 1 324...
Page 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Page 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Page 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Page 337: ...T rk e 337...
Page 358: ...358 Slovensk jazyk...
Page 368: ...368 Slovensk jazyk...
Page 379: ...Slovenski 379...
Page 400: ...400 Hrvatski...
Page 417: ...1 2 PVC 1 2 417...
Page 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Page 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Page 420: ...MSDS M48 M48 O ring 420...
Page 421: ...421...
Page 422: ...6 12 435 422...
Page 423: ...423...
Page 432: ...1 1 432...
Page 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Page 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Page 440: ......
Page 441: ......