Recalibrate the sensor whenever disconnected from power and removed
from water for more than 15 minutes.
Choosing the calibration method
An initial 2-point calibration, including both zero
3
and slope
(process concentration) measurements, must be done for new or
refurbished sensors.
One or two measurements are required for sensor calibration.
Measurements are made with the chlorine sensor in the flow cell.
When only one measurement is made (1-point calibration), a zero
calibration or a process concentration measurement (grab sample
analysis) is done. A zero calibration can be done chemically by
measuring water with no chlorine or electrically by electronically
removing the calibration offset produced by the gateway
3
. A process
concentration measurement is done chemically by measuring the
chlorine concentration of the process flow with a reference method (grab
sample analysis) and then entering the measured value through the
controller.
When two measurements are made (2-point calibration), the first data
point measured is the zero-point and is determined by doing a zero
calibration either electrically or chemically using the same method as the
1-point calibration
3
. The second data point measured is the process
concentration and is determined by doing a process concentration
measurement (grab sample analysis) using the same method as the 1-
point calibration.
Note: When a chemical zero calibration is done, the instrument automatically does
an electrical zero calibration before the chemical zero calibration and shows both
offsets in the calibration results.
1-point calibration procedure
1.
Push the
MENU
key and select Sensor Setup, Calibrate, Chlorine.
2.
Select 1 Point Sample.
3.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
4.
Select either Zero Cal (zero calibration) or Process Conc (process
concentration calibration).
5.
Select the option for the output signal during calibration:
Option
Description
ACTIVE
The instrument sends the current measured output value
during the calibration procedure.
HOLD
The sensor output value is held at the current measured value
during the calibration procedure.
TRANSFER
A preset output value is sent during calibration. Refer to the
controller user manual to change the preset value.
6.
If Process Conc was selected:
a.
Push
ENTER
.
The measured value is shown.
b.
Wait for the value to stabilize and push
ENTER
4
to enter a
process flow measurement.
c.
Measure the chlorine concentration of the process flow (grab
sample analysis) with an instrument using a reference method
(e.g., DPD). Use the arrow keys to enter the measured value and
push
ENTER
.
7.
If Zero Cal was selected, select the type of calibration:
Option
Description
ELECTRICAL
The offset produced by the gateway is removed to set the
zero-point (no sample used).
CHEMICAL
Water with no chlorine is measured to set the zero-point.
8.
If Electrical was selected, wait for the value to stabilize and push
ENTER
4
to set the zero-point electrically.
3
An electrical zero calibration is recommended. A chemical zero calibration is only recommended if the process flow chlorine concentrations is
routinely lower than 0.5 ppm.
4
If the option for Auto Stab (auto stabilization) is set to Yes in the Calibration Options menu, the screen will advance to the next step automatically.
Refer to
Change the calibration options
on page 15.
English
13
Summary of Contents for Chlorine Sensor
Page 8: ...8 English...
Page 17: ...English 17...
Page 29: ...Deutsch 29...
Page 51: ...Assemblare il sensore per cloro Italiano 51...
Page 53: ...Italiano 53...
Page 73: ...Assemblage du capteur de chlore Fran ais 73...
Page 75: ...Fran ais 75...
Page 95: ...Montaje del sensor de cloro Espa ol 95...
Page 97: ...Espa ol 97...
Page 105: ...Espa ol 105...
Page 117: ...Montar o sensor de cloro Portugu s 117...
Page 119: ...Portugu s 119...
Page 139: ...e tina 139...
Page 160: ...160 Dansk...
Page 181: ...Chloorsensor samenbouwen Nederlands 181...
Page 183: ...Nederlands 183...
Page 203: ...Monta czujnika chloru Polski 203...
Page 205: ...Polski 205...
Page 225: ...Svenska 225...
Page 235: ...Svenska 235...
Page 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Page 244: ...1 1 4 2 5 3 6 30 mV 244...
Page 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Page 246: ...48 246...
Page 247: ...247...
Page 248: ...6 12 260 248...
Page 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Page 257: ...6 1 3 6 3 1 1 6 257...
Page 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Page 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Page 270: ...270 Magyar...
Page 280: ...280 Magyar...
Page 291: ...Asamblarea senzorului de clor Rom n 291...
Page 293: ...Rom n 293...
Page 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Page 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Page 311: ...OFF 30 pH pH 2 311...
Page 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Page 313: ...313...
Page 314: ...6 12 327 314...
Page 315: ...315...
Page 324: ...1 1 324...
Page 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Page 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Page 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Page 337: ...T rk e 337...
Page 358: ...358 Slovensk jazyk...
Page 368: ...368 Slovensk jazyk...
Page 379: ...Slovenski 379...
Page 400: ...400 Hrvatski...
Page 417: ...1 2 PVC 1 2 417...
Page 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Page 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Page 420: ...MSDS M48 M48 O ring 420...
Page 421: ...421...
Page 422: ...6 12 435 422...
Page 423: ...423...
Page 432: ...1 1 432...
Page 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Page 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Page 440: ......
Page 441: ......