Specifications
Specifications are subject to change without notice.
Specification
Details
Measurement range
0 to 20 ppm
1
Lower limit of detection
30 ppb (0.030 ppm)
Resolution
0.001 ppm (1 ppb)
pH operating range
4 to 9 pH units
Accuracy
(chlorine concentrations
within ±2 ppm or 20%
(whichever is smaller) of the
calibrated point)
Free chlorine (0 to 10 ppm):
• ± 3% of the reference test
2
(DPD) at constant
pH less than 7.2 (±0.2 pH unit)
• ± 10% of the reference test
2
(DPD) at stable
pH less than 8.5 (±0.5 pH unit from the pH at
chlorine calibration)
Total chlorine (0 to 10 ppm):
• ± 10% of the reference test
2
(DPD) at stable
pH less than 8.5 (±0.5 pH unit from the pH at
chlorine calibration)
• ± 20% of the reference test
2
(DPD) at pH
greater than 8.5
Specification
Details
Repeatability
30 ppb or 3%, whichever is greater
Response time
Free chlorine: 140 seconds for 90% change (T
90
)
at a stable temperature and pH
Total chlorine: 100 seconds for 90% change (T
90
)
at a stable temperature and pH
Sampling time
Continuous
Interferences
Free chlorine: Monochloramine, chlorine dioxide,
ozone and chalk deposits
Total chlorine: Chlorine dioxide, ozone and chalk
deposits
Pressure limit
0.5 bar, no pressure impulses and/or vibrations
Flow rate
30 to 50 L/hour (7.9 to 13.2 gal/hour)
40 L/hour (10.5 gal/hour) – optimal
Power requirements
12 VDC, 30 mA maximum (supplied by controller)
Operating temperature
5 to 45 °C (41 to 113 °F)
Storage temperature
Sensor: 0 to 50 °C (32 to 122 °F) dry without
electrolyte
Electrolyte: 15 to 25 °C (59 to 77 °F)
Dimensions
(length/diameter)
195 mm (7.68 in.)/25 mm (0.98 in.)
Cable length/type
1 m
Cable connection
5 pin, M12 connector
Measurement method
Reagentless, electrochemical, amperometric,
three-electrode system
Calibration methods
1-point or 2-point (zero and slope) calibration
Temperature compensation
Internal temperature sensor
1
The chlorine sensors are not applicable for applications with consistently low (< 0.1 ppm) or zero chlorine concentrations.
2
The test must be conducted at the analyzer sampling point.
English
3
Summary of Contents for Chlorine Sensor
Page 8: ...8 English...
Page 17: ...English 17...
Page 29: ...Deutsch 29...
Page 51: ...Assemblare il sensore per cloro Italiano 51...
Page 53: ...Italiano 53...
Page 73: ...Assemblage du capteur de chlore Fran ais 73...
Page 75: ...Fran ais 75...
Page 95: ...Montaje del sensor de cloro Espa ol 95...
Page 97: ...Espa ol 97...
Page 105: ...Espa ol 105...
Page 117: ...Montar o sensor de cloro Portugu s 117...
Page 119: ...Portugu s 119...
Page 139: ...e tina 139...
Page 160: ...160 Dansk...
Page 181: ...Chloorsensor samenbouwen Nederlands 181...
Page 183: ...Nederlands 183...
Page 203: ...Monta czujnika chloru Polski 203...
Page 205: ...Polski 205...
Page 225: ...Svenska 225...
Page 235: ...Svenska 235...
Page 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Page 244: ...1 1 4 2 5 3 6 30 mV 244...
Page 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Page 246: ...48 246...
Page 247: ...247...
Page 248: ...6 12 260 248...
Page 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Page 257: ...6 1 3 6 3 1 1 6 257...
Page 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Page 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Page 270: ...270 Magyar...
Page 280: ...280 Magyar...
Page 291: ...Asamblarea senzorului de clor Rom n 291...
Page 293: ...Rom n 293...
Page 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Page 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Page 311: ...OFF 30 pH pH 2 311...
Page 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Page 313: ...313...
Page 314: ...6 12 327 314...
Page 315: ...315...
Page 324: ...1 1 324...
Page 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Page 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Page 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Page 337: ...T rk e 337...
Page 358: ...358 Slovensk jazyk...
Page 368: ...368 Slovensk jazyk...
Page 379: ...Slovenski 379...
Page 400: ...400 Hrvatski...
Page 417: ...1 2 PVC 1 2 417...
Page 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Page 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Page 420: ...MSDS M48 M48 O ring 420...
Page 421: ...421...
Page 422: ...6 12 435 422...
Page 423: ...423...
Page 432: ...1 1 432...
Page 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Page 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Page 440: ......
Page 441: ......