7.
If Chemical was selected:
a.
Turn off the process flow and put water with no chlorine through
the flow cell. Make sure the temperature of the water with no
chlorine is as close as possible to the process sample flow
temperature.
b.
Push
ENTER
.
The measured value is shown.
c.
Wait for the value to stabilize and push
ENTER
5
to set the zero-
point chemically.
d.
Turn off the flow of water with no chlorine and turn on the process
flow.
e.
Push
ENTER
.
The measured value is shown.
f.
Wait for the value to stabilize and push
ENTER
5
to enter a
process flow measurement.
8.
Measure the chlorine concentration of the process flow (grab sample
analysis) with an instrument using a reference method (e.g., DPD).
Use the arrow keys to enter the measured value and push
ENTER
.
9.
Review the calibration result:
• Passed—the sensor is calibrated and ready to measure samples.
The slope and/or offset values are shown.
• Failed—the calibration slope or offset is outside of accepted limits.
Do sensor maintenance (refer to
on page 16) and
then repeat the calibration.
10.
If the calibration passed, push
ENTER
to continue.
11.
If the option for operator ID is set to Yes in the Calibration Options
menu, enter an operator ID. Refer to
Change the calibration options
on page 15.
12.
On the New Sensor screen, select whether the sensor is new:
Option Description
YES
The sensor was not calibrated previously with this instrument. The
days of operation and previous calibration curves for the sensor are
reset.
NO
The sensor was calibrated previously with this instrument.
13.
Return the sensor to the process and push
ENTER
.
The output signal returns to the active state and the measured
sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when
the outputs return to the active state.
Reset calibration to defaults
To remove a bad calibration, replace the user calibration settings with
the default calibration settings using the Calibrate menu. Then
recalibrate the sensor when needed.
1.
Push the
MENU
key and select Sensor Setup, Calibrate, [Select
Sensor], Reset Defaults.
2.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
3.
Select Yes and push
Enter
.
Change the calibration options
The user can set a calibration reminder, enable auto stabilization during
calibrations or include an operator ID with calibration data from the Cal
Options menu.
1.
Push the
MENU
key and select Sensor Setup, Calibrate, [Select
Sensor], Cal Options.
2.
Customize the options:
Option
Description
AUTO STAB
Enables the system to accept measurement signal values
during calibrations and advance to the next step of the
calibration when the system determines the measurement
signal has stabilized-On or Off (default). Enter a
stabilization range-25 to 75 ppb (0.025 to 0.075 ppm).
CAL REMINDER
Sets a reminder for the next calibration in days, months or
years.
OP ID ON CAL
Includes an operator ID with calibration data—Yes or No
(default). The ID is entered during the calibration.
English
15
Summary of Contents for Chlorine Sensor
Page 8: ...8 English...
Page 17: ...English 17...
Page 29: ...Deutsch 29...
Page 51: ...Assemblare il sensore per cloro Italiano 51...
Page 53: ...Italiano 53...
Page 73: ...Assemblage du capteur de chlore Fran ais 73...
Page 75: ...Fran ais 75...
Page 95: ...Montaje del sensor de cloro Espa ol 95...
Page 97: ...Espa ol 97...
Page 105: ...Espa ol 105...
Page 117: ...Montar o sensor de cloro Portugu s 117...
Page 119: ...Portugu s 119...
Page 139: ...e tina 139...
Page 160: ...160 Dansk...
Page 181: ...Chloorsensor samenbouwen Nederlands 181...
Page 183: ...Nederlands 183...
Page 203: ...Monta czujnika chloru Polski 203...
Page 205: ...Polski 205...
Page 225: ...Svenska 225...
Page 235: ...Svenska 235...
Page 243: ...ESD 10 pH CLF10sc CLT10sc sc 1 243...
Page 244: ...1 1 4 2 5 3 6 30 mV 244...
Page 245: ...pH 2 2 1 4 2 5 3 O MSDS 48 245...
Page 246: ...48 246...
Page 247: ...247...
Page 248: ...6 12 260 248...
Page 249: ...0 1 ppm mg L Configure 1 MENU Sensor Setup Calibrate Cal Options EDIT NAME 10 249...
Page 257: ...6 1 3 6 3 1 1 6 257...
Page 258: ...1 2 3 245 4 M48 5 6 7 3 8 9 10 6 12 260 11 258...
Page 259: ...3 1 3 2 4 1 2 3 245 4 5 6 7 8 6 12 260 9 1 2 3 245 4 M48 5 6 1 2 7 8 9 260 10 259...
Page 270: ...270 Magyar...
Page 280: ...280 Magyar...
Page 291: ...Asamblarea senzorului de clor Rom n 291...
Page 293: ...Rom n 293...
Page 309: ...12 30 5 45 C 41 113 F 0 50 C 32 122 F 15 25 C 59 77 F 195 7 68 25 0 98 1 5 M12 1 2 1 2 309...
Page 310: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 310...
Page 311: ...OFF 30 pH pH 2 311...
Page 312: ...2 1 4 2 5 3 MSDS M48 M48 312...
Page 313: ...313...
Page 314: ...6 12 327 314...
Page 315: ...315...
Page 324: ...1 1 324...
Page 325: ...1 2 3 312 4 M48 5 6 7 3 8 9 10 6 12 327 11 325...
Page 326: ...3 1 3 2 4 1 2 3 312 4 5 6 7 8 6 12 327 9 1 2 3 312 4 M48 5 6 1 2 7 8 9 327 10 326...
Page 327: ...MSDS 3 308 1 2 3 4 1 312 2 3 4 5 6 7 6 12 327 8 1 1 GATEWAY INFO RST CAL HISTORY SIGNALS 327...
Page 337: ...T rk e 337...
Page 358: ...358 Slovensk jazyk...
Page 368: ...368 Slovensk jazyk...
Page 379: ...Slovenski 379...
Page 400: ...400 Hrvatski...
Page 417: ...1 2 PVC 1 2 417...
Page 418: ...pH 10 pH CLF10sc CLT10sc sc 1 1 1 4 2 5 3 6 LED LED LED 418...
Page 419: ...LED LED 30 mV pH pH 2 2 1 4 2 5 3 O ring 419...
Page 420: ...MSDS M48 M48 O ring 420...
Page 421: ...421...
Page 422: ...6 12 435 422...
Page 423: ...423...
Page 432: ...1 1 432...
Page 433: ...1 2 3 420 4 M48 5 6 7 3 8 9 10 6 12 435 11 433...
Page 434: ...3 1 3 2 4 1 2 3 420 4 5 6 7 8 6 12 435 9 1 2 3 420 4 M48 5 6 1 2 7 8 9 435 10 434...
Page 440: ......
Page 441: ......