Guidance Tables
Table 201 – Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Medical Electrical Equipment and Medical Electrical Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Visutron 900
Touch
is intended for use in the electromagnetic environment specified below. The
customer or user of the Visutron 900
Touch
should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment
- Guidance -
Conducted and Radiated
RF Emissions
CISPR 11
Group 1
The Visutron 900
Touch
uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Conducted and Radiated
RF Emissions
CISPR 11
Class A
The Visutron 900
Touch
is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies building
for domestic power.*
Harmonic Distortion
IEC 61000-3-2
Class A
Voltage Fluctuations and
Flicker
IEC 61000-3-3
Complies
Note: *The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class
B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment
Bluetooth RF Transmitter Characteristics - Central Unit
• Contains FCC ID: QOQWT11u
• Contains IC: 5123A - 5123A-WT11u
• Frequency 2402-2480 MHz, Spread Spectrum, 0.045 W
Bluetooth Adapter (Dongle) Characteristics
•
Contains FCC ID: QOQWT11μ
•
Contains IC: 5123A-5123A-WT11μ
• Frequency 2402-2480 MHz, Spread Spectrum, 0.045 W
FCC / IC Statements
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause undesired
operation.
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.
End users must follow the specific operating instructions for satisfying RF exposure compliance. This
transmitter must not be co-located or operating in conjunction with any other antenna or transmitter, ex-
cept in accordance with FCC test procedures. This transmitter is considered as mobile device.
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the
following two conditions: (1) this device may not cause interference, and (2) this device must accept any
interference, including interference that may cause undesired operation of the device.
The minimum separation distance to human body 40 mm. RF exposure or SAR evaluations is not required
when the separation distance is 40 mm or more.