Press the yellow
C
button (in the upper left corner) or tap
Clear
on the bottom of the controller screen to
enter Clear mode.
Note:
Clear Mode will be active until the
ENTER
button is pressed, or
Exit
is tapped, to exit Clear Mode.
Figure PE-15, Clear Mode
Clearing a Single Field
Tap the box on the screen for the individual data field to be cleared (
e.g.,
Tap the Right Eye, Sphere).
Clearing Data for One or Both Eyes
Press the
R
,
L
,
or
B
button, or tap
R
,
L
, or
B
next to the lens apertures.
Clearing Individual Data Sets
Tap
Subj
,
Final
,
AR
,
LM
,
REF
,
Mem 1
, or
Mem 2
, located on the menu bar at the bottom of the screen,
or press the corresponding grey button, to indicate the data set to be cleared.
Clearing All Data
Press the
C
button, or tap
Clear
, to clear all the data. This will reset the entire exam and prepare the
Visutron 900
Touch
for the next patient.
Reset Head
Tap
Reset Head
to reinitialize the Phoroptor head.
Performing an Exam
(continued)
67
16243-101 Rev. G