Connecting the Visutron 900
Touch
System
(continued)
1. Align the pins in the Phoroptor Cable with the Phoroptor head.
Refer to figure IS-03.
2. Connect the cable by turning the metal collar clockwise until it is
fully secure. Refer to figure IS-04.
3. Connect the other end of the Phoroptor Cable to the Phoroptor
port in the Central Unit. Refer to Figure IS-05.
4. Connect the built-in cable on the Controller to the Controller port
on the Central Unit.
5. Connect the Power Cord to the Central Unit, then plug in the
power cord to an outlet providing appropriate voltage. Turn the
unit on by turning the ON/OFF switch to ON.
6. Connect the Projector to the Projector port on the Central Unit in
one of the following ways.
a. Refer to the
Connecting with Cables
section if connecting
with cable.
b. Refer to the
Connecting with Bluetooth Dongles
section if
connecting with Bluetooth.
c. Refer to the
Connecting With Bluetooth USB (ClearChart 4
Series Only)
section if connecting with Bluetooth USB.
7. Connect Lensmeters or Auto Refractors to the Instrument 1 or
Instrument 2 ports on the Central Unit in one of the following
ways.
a. Refer to the
Connecting with Cables
section if connecting
with cable.
b. Refer to the
Connecting with Bluetooth Dongles
section if
connecting with Bluetooth.
8. Connect a computer to the EMR port on the Central Unit in one
of the following ways. Refer to the
EMR Setup
section for setup
instructions.
a. Refer to the
Connecting with Cables
section if connecting
with cable.
b. Refer to the
Connecting with Bluetooth Dongles
section if
connecting with Bluetooth.
9. If needed, connect the Transfer port to another Visutron 900
Touch
in one of the following ways to send data from one device
to another.
Refer to the
Daisy Chaining Multiple Visutron 900
Touch’s
section for instructions.
a. Refer to the
Connecting with Cables
section if connecting
with cable.
b. Refer to the
Connecting with Bluetooth Dongles
section if
connecting with Bluetooth.
c. Refer to the
Connecting with Internal Bluetooth
section if
connecting with Bluetooth.
Connecting a Right or Left Mount Visutron 900
Touch
Connect the Visutron 900 Touch using the connection cable (559-276)
according to Figure IS-05A.
Instrument Setup
(continued)
Figure IS-03, Phoroptor Cable
Figure IS-04, Phoroptor Connected
Figure IS-05, Phoroptor Connected
Figure IS-05A, Connection Diagram
12
16243-101 Rev. G
Summary of Contents for VISUTRON 900 Touch
Page 1: ...VISUTRON 900 Touch User s Guide ...
Page 85: ...Notes 85 16243 101 Rev G ...
Page 86: ...Notes 86 16243 101 Rev G ...
Page 87: ...Notes 87 16243 101 Rev G ...