PASS-FAIL MODE
79
Set Hi limit
Move the cursor to Hi column.
Enter the Hi limit using the
numerical keys and unit keys.
Spd
Hi
1.0000H
0.0000H
Lo
MAX
Range
follows the specification for each
measurement item
Example: 1.5kH
(for Ls)
Enter
H
k
Set Lo limit
Move the cursor to Lo column.
Enter the Lo limit using the
numerical keys and unit keys.
Spd
Hi
0.0000H
Lo
MAX
1.0000H
Range
follows the specification for each
measurement item
Example: 1.0kH
(for Ls)
Enter
H
k
Set trigger delay
Move the cursor to Dly (Delay)
column. Enter the amount of
trigger delay time using the
numerical keys and unit keys.
Hi
Lo
10 ms
Dly
1.0000H
0.0000H
Range
0 (no delay) ~ 1000ms
Example: 10ms
m