LCR-8000G Series User Manual
120
:graph:freq?
Returns the frequency if the sweep mode is drive
level
:graph:lev <NR3>
Set the drive level if the sweep mode is frequency.
Parameter: (.1 volts) .1v, 100m, 1e-1, 1.000000e-1
Example: :graph:lev 100m
NOTE: e1 or e+1 is invalid for the lev
command. 2 volts maximum.
:graph:lev?
Returns the drive level if the sweep mode is
frequency.
:graph:speed
<disc>
Set measurement speed for the sweep.
Parameter: fast, med, slow
Example: :graph:speed med
:graph:speed?
Returns the measurement speed of the sweep.
Return Parameter: 1(fast), 2 (med), 3(slow)
:graph:step
<NR1>
Select the number of pixels between each
measured point.
Parameter: 1(step size 1),2(step size 2 ),3(step size
4),4(step size 8)
Example: (step size 8) :graph:step 4
:graph:step?
Query the current step size for the plot.
:graph:hi-lim
<NR3>
Set the maximum value for Y-axis in the graph
mode.
Parameter: real number up to 1^12 (1e+12)
Example: graph:hi-lim 8.5e9
Note: Set the low limit before setting the high limit.
The graph limits will only work whilst the
“autofit” function is set to “off”
:graph:hi-lim?
Returns the maximum value of Y-axis in the graph
mode.
www.
.com
1.800.561.8187