LCR-8000G Series User Manual
6
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
Safety Guidelines
General
Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at circuits directly
connected to Mains (Note below).
Do not disassemble the instrument unless you
are qualified as service personnel.
(Measurement categories) EN 61010-1:2001 specifies the
measurement categories and their requirements as follows. the
LCR-8000G falls under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage: 115V (+10% / -25%), AC
230V (+15% / -14%) (Selectable), 50/60Hz.
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid
electrical shock.