LCR-8000G Series User Manual
138
Qx
Measured Q value
De
Relative D accuracy
[1]High impedance threshold recommended greater than 1k
Ω.
[2]Low impedance threshold recommended less than 1k
Ω.
Specification
Test Frequency
LCR-8101G
DC, AC:20Hz~1MHz
LCR-8105G
DC, AC:20Hz~5MHz
LCR-8110G
DC, AC:20Hz~10MHz
Basic Accuracy
R,Z,X,G,Y,B,L,C ±0.1%@1kHz
R&G Accuracy
When Qx
≧
0.1, multiply Ae by
2
Q
1
for R, G
accuracies.
Measurable
Parameters
Rac, Rdc, Rs, Rp, Z, Ls, Lp, D, G, B,
θ
, Cp, Cs, Q, Y, X
Measurement
Range
R,Z,X
0.1m
Ω
~100M
Ω
G, Y, B
10ns~1ks
L
0.1nH~100kH
C
0.01pF~1F
D
0.00001~1000
Q
0.01~9999.9
Rdc
0.01m
Ω
~100M
Ω
Ɵ
-180°~+180°
Equivalent
Circuit
Parallel
C+G, L+G, B+G, B+Q, B+D, B+R
Series
X+R, X+D, X+Q
Series & Parallel C+R, C+D, C+Q, L+R, L+D, L+Q
Polar Form
Z + Phase Angle, Y + Phase Angle
Input Impedance 100
Ω
Measurement
Speed
DC
Max:30ms; Fast:60ms;
Medium:120ms; Slow:900ms
AC
≤
100Hz
Max:600ms; Fast:650ms;
Medium:1.2s; Slow:1.3s
AC
≤
2kHz
Max:120ms; Fast:180ms;
Medium:470ms; Slow:600ms
AC
>
2kHz
Max:75ms; Fast:150ms;
Medium:450ms; Slow:600ms
AC=1MHz
(LCR-8101G)
Max:120ms; Fast:150ms;
Medium:470ms; Slow:620ms
www.
.com