LCR-8000G Series User Manual
30
Pass/Fail test (Multiple step)
Step
Description
Details
1. Connect fixture
Connect the fixture to the DUT.
Page24
2. Set buzzer
sound
Press the Menu key, then F5 (System).
Press the Up/Down arrow key to move
the cursor to Beep, then use the
Left/Right key to select setting (Off
recommended).
Page73
3. Set Average
Press the Menu key, then F5 (System).
Press the Up/Down arrow key to move
the cursor to Average, then use the
numerical keys to enter an average
number (1-256). Press enter to confirm the
number.
Page74
4. Enter multiple
step mode
Press the Menu key, then F3 (Multi step). Page75
5. Select
measurement
item
Press the arrow keys to move the cursor
to Step 01 Func. Press F1 (Prog)
repeatedly to select the item.
Page77
6a. Set
parameters
Press the arrow keys to move the cursor
to the parameters below. Use the
numerical and unit keys for editing values
or F1 (Prog) for selecting options.
Page77
6b. Add steps
Move the cursor to the first empty step
and press F1 (Prog).
Page77
6c. Copy to the
next step
Press F2 (Copy). The selected step
contents are copied and inserted to the
next step.
Page80
6d. Delete step
Press F3 (Delete). The selected step is
deleted.
Page80
7. Save program
Press F4 (Save). The edited program is
saved.
Page84
www.
.com
1.800.561.8187