REMOTE CONTROL
115
:meas:func:minor
?
Returns the second AC function.
Parameter: 0 (Q), 1 (D), 2 (R), 3 (G)
If the first function is Z or Y, this command returns
the last non-polar setting
:meas:hi-lim
<NR2>
:meas:hi-lim?
Sets or returns scale high limit as percentage.
Example: :meas:hi-lim 5.0 (+5.0%)
:meas:lev <NR2>
:meas:lev?
Sets or returns drive level for currently selected
test.
Parameter example: (200mV) 0.2V, 200m
:meas:limit
<disc>
:meas:limit?
Sets or returns percentage, absolute or delta scale
limits.
Send parameter: abs (absolute), perc (percentage),
delta (delta)
Return parameter: 0 (absolute), 1 (percentage), 2
(delta)
:meas:lo-lim
<NR2>
:meas:lo-lim?
Sets or returns scale low limit as percentage.
Example: :meas:hi-lim –5.0 (–5.0%)
:meas:nom
<NR3>
:meas:nom?
Sets or returns nominal value for scale.
Send parameter: according to the active unit (1e-6f
= 1uF)
Return parameter example: .10000000e-1 = 10mH
:meas:range
<NR1>
:meas:range?
Selects or returns auto-ranging or range-hold on
range N.
Send parameter: auto, hold, 1 ~ 7
Return parameter: 0 (auto), 1 ~ 7
:meas:scale
<disc>
:meas:scale?
Shows or hides the scale bar or returns the status.
Send parameter: on, off
Return parameter: 0 (scale hidden), 1 (scale visible)