GDS-1000B Programming Manual
200
Related
commands
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:MODe
Parameter
<sting>
7/10 characters, must be enclosed in
double quotes "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the address value in binary.
Example 1
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:MODe
ADDR7
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:VALue
"xxx0101"
Sets the address to XXX0101
Example 2
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:VALue?
XXX0101
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess
:DIRection
Description
Sets or queries the address bit as read write or
don’t care for the search function.
Note
This setting only applies when the I
2
C search
trigger is set to trigger on Address or
Address/Data
Syntax
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:DIRection
{ READ | WRITE | NOCARE | ? }
Related
commands
:SEARCH:TRIGger:BUS:B1:I2C:CONDition
Parameter
READ
Set read as the data direction.
WRITE
Set write as the data direction.
NOCARE
Set either as the data direction.
Return Parameter
Returns the direction (READ, WRITE, NOCARE).
Set