GDS-1000B Programming Manual
160
Example
:TRIGger:BUS:B1:LIN:DATa:SIZe?
>1
:TRIGger:BUS:B1:LIN:DATa:SIZe 2
:TRIGger:BUS:B1:LIN:DATa:SIZe?
>2
:TRIGger:BUS:B1:LIN:DATa:VALue
Description
Sets or returns the binary data string to be used for
the LIN trigger.
Note: Only applicable when the condition is set to
DATA or IDANDDATA.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Related
Commands
:TRIGger:BUS:B1:LIN:DATa:SIZe
Syntax
:TRIGger:BUS:B1:LIN:DATa:VALue {<string>|?}
Parameter/
Return parameter
<string>
The size of the string depends on the
data size setting. The string must be
enclosed in double quotes, “string”.
String contents:
x = don’t care
1 = binary 1
0 = binary 0
Example
:TRIGger:BUS:B1:LIN:DATa:SIZe 1
:TRIGger:BUS:B1:LIN:DATa:VALue "01010X1X"
:TRIGger:BUS:B1:LIN:DATa:VALue?
>01010X1X
Set
Query