
GDS-1000B Programming Manual
204
Example
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:SIZe 5
Sets the number of bytes to 5.
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:
VALue
Description
Sets or queries the search triggering data value for
the UART bus when the bus is set to trigger on Rx
Data.
Syntax
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:VALue
{<string> | ? }
Related
commands
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:SIZe
Parameter
<sting>
The number of characters in the string
depends on the data size setting. The
string must be enclosed in double
quotes, "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the data value.
Example1
:SEARCH:TRIGger:BUS:B1:UART:CONDition RXDATA
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:SIZe 1
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:VALue
"1x1x0101"
Sets the value to 1x1x0101
Example 2
:SEARCH:TRIGger:BUS:B1:UART:RX:DATa:VALue?
1X1X0101
:SEARCH:TRIGger:BUS:B1:UART:TX:DATa
:SIZe
Description
Sets or queries the number of bytes for UART data.
Set
Query
Set
Query