COMMAND DETAILS
231
Parameter/
Return parameter
<NR1>
1~20
Example1
:SEGMents:MEASure:PLOT:DIVide 5
Sets the number of bins to 5 for the statistics function.
:SEGMents:MEASure:PLOT:SELect
Description
Sets or queries which bin to view the statics of.
Syntax
:SEGMents:MEASure:PLOT:SELect {<NR1> | ? }
Related
commands
:SEGMents:MEASure:PLOT:SOURce
:SEGMents:MEASure:PLOT:DIVide
Parameter
<NR1>
1~20
(cannot exceed the number of bins)
Return parameter Return the bin number as <NR3>.
Example1
:SEGMents:MEASure:PLOT:SELect 5
Set to bin number 5.
:SEGMents:MEASure:PLOT:RESults
Description
Returns the results of the currently selected bin for
the statistics measurement.
Note
At least one automatic measurement must be
turned on.
Syntax
:SEGMents:MEASure:PLOT:RESults?
Related
commands
:SEGMents:STATE
:SEGMents:MEASure:MODe PLOT
:SEGMents:MEASure:PLOT:SOURce
:SEGMents:MEASure:PLOT:DIVide
:SEGMents:MEASure:PLOT:SELect
Return parameter
Returns the statistics measurements as a string.
Set
Query