GDS-1000B Programming Manual
136
Parameter
MORE than >
LESSthan
<
EQual
=
UNEQual
≠
Return parameter
Returns the pulse width trigger conditions.
Example
:TRIGger:PULSe:WHEn UNEQual
Sets the trigger pulse width conditions to not equal
to (≠).
:TRIGger:PULSe:TIMe
Description
Sets or queries the pulse width time.
Syntax
:TRIGger:PULSe:TIMe {<NRf> | ?}
Related
commands
:TRIGger:TYPe
:TRIGger:PULSe:WHEn
Parameter
<NRf>
Pulse width time (4ns~10s)
Return parameter <NR3>
Returns the pulse width time in
seconds.
Example
:TRIGger:PULSe:TIMe 4.00E-5
Sets the trigger pulse width to 40.0us.
:TRIGger:TIMEOut:WHEn
Description
Sets or queries the timeout trigger condition.
Syntax
:TRIGger:TIMEOut:WHEn {HIGH|LOW|EITher|?}
Related
commands
:TRIGger:TIMEOut:TIMER
Parameter
HIGH
Signal is high.
LOW
Signal is low.
EITher
Signal is high or low.
Set
Query
Set
Query