GDS-1000B Programming Manual
74
Display Commands
:DISPlay:INTensity:WAVEform .................................... 74
:DISPlay:INTensity:GRATicule .................................... 74
:DISPlay:INTensity:BACKLight .................................... 75
:DISPlay:INTensity:BACKLight:AUTODim:ENAble .... 75
DISplay:INTENSITy:BACKLight:AUTODim:TIMe ...... 75
:DISPlay:PERSistence .................................................. 76
:DISPlay:GRATicule ..................................................... 76
:DISPlay:WAVEform .................................................... 77
:DISPlay:OUTPut ......................................................... 77
:DISPlay:INTensity:WAVEform
Description
Sets or queries the waveform intensity level.
Syntax
:DISPlay:INTensity:WAVEform {<NRf> | ?}
Parameter
<NRf>
0.0E+0 to 1.0E+2 (0% to 100%)
Return Parameter <NR3> Returns the intensity.
Example
:DISPlay:INTensity:WAVEform 5.0E+1
Sets the waveform intensity to 50%.
:DISPlay:INTensity:GRATicule
Description
Sets or queries the graticule intensity level.
Syntax
:DISPlay:INTensity:GRATicule {<NRf> | ?}
Parameter
<NRf>
1.0E+0 to 1.0E+2 (10% to 100%)
Return Parameter <NR3> Returns the graticule intensity.
Example
:DISPlay:INTensity:GRATicule 5.0E+1
Sets the graticule intensity to 50%.