![GW Instek GDS-1000B Series Programming Manual Download Page 152](http://html1.mh-extra.com/html/gw-instek/gds-1000b-series/gds-1000b-series_programming-manual_2280038152.webp)
GDS-1000B Programming Manual
152
Example2
:TRIGger:BUS:B1:SPI:DATa:MOSI:VALue?
1X1X0101
:TRIGger:BUS:B1:CAN:CONDition
Description
Sets or returns the CAN trigger condition.
Note
This command is only applicable when the Serial
Bus Decode (DS1B-BUS) option is installed. See the
user manual for details.
Syntax
:TRIGger:BUS:B1:CAN:CONDition
{SOF|FRAMEtype|IDentifier|DATA|IDANDDATA|EOF|
ACKMISS|STUFFERR|?}
Parameter/
Return parameter
SOF
Triggers on a start of frame
FRAMEtype
Triggers on the type of frame
Identifier
Triggers on a matching identifier
DATA
Triggers on matching data
IDANDDATA
Triggers on matching identifier and
data field
EOF
Triggers on the end of frame
ACKMISS
Triggers on a missing acknowledge
STUFFERR
Triggers on a bit stuffing error
Example1
:TRIGger:BUS:B1:CAN:CONDition SOF
Triggers on a start of frame.
Example2
:TRIGger:BUS:B1:CAN:CONDition?
>SOF
:TRIGger:BUS:B1:CAN:FRAMEtype
Description
Sets or returns the frame type for a CAN
FRAMEType trigger.
Set
Query
Set
Query