Example
: BUS1:DISplay:FORMAt BINary
Sets the display format to binary.
:LISTer:DATA
Description
Returns the Event Table data as a binary block
data.
Syntax
:LISTer:DATA?
Return Parameter Returns the event table as binary block data. The
binary block data contains comma separated data with
new lines at the end of each row.
:BUS1:CAN:SOURce
Description
Sets or returns the CAN input source.
Syntax
:BUS1:CAN:SOURce { CH1 | CH2 | CH3 | CH4 | ? }
Parameter/Return
parameter
CH1 ~ CH4 Analog channel source
Example
:BUS1:CAN:SOURCE?
>CH1
Returns the CAN source.
:BUS1:CAN:PROBe
Description
Sets or returns the signal type of the CAN bus.
Syntax
:BUS1:CAN:PROBe {CANH | CANL | TX | RX | ? }
Parameter/Return
parameter
CANH
CAN-High
CANL
CAN-Low
TX
Transmit
RX
Receive
Query
Set
Query
Set
Query