COMMAND OVERVIEW
155
Example
->:CALC:STAT:RES:MIN?
->+354.33E+0,7
:CALCulate:STATistics:RESistance:LIMit?
Description:
Queries the result count of statistics for resistance.
Query Syntax
:CALCulate:STATistics:RESistance:LIMit?
Return parameter <HI count>, <OK count>, <LO count>, <FAULT
count>
Example
->:CALC:STAT:RES:LIM?
->0, 10, 0, 0
Note
When querying the file count of comparator, make
sure the comparator function is enabled, otherwise 0,
0, 0, 0 will be returned for data.
:CALCulate:STATistics:RESistance:DEViation?
Description:
Queries the standard deviation value of statistics
for resistance.
Query Syntax
:CALCulate:STATistics:RESistance:DEViation?
Return parameter < Standard deviation of maternal
σ
n
>,
<Standard deviation of samples
σ
n-1
>
Example
->:CALC:STAT:RES:DEV?
->0.0016, 0.0017
Note
Standard deviation of maternal:
Standard deviation of samples:
Set
Query
Set
Query