GBM-3080/3300 User Manual
116
Note
If the host can’t accept the data returned by the
device, you can use the following methods to try to
solve:
The hand shake function is disabled. Please enable
the hand shake function. Refer to the text below for
hand shake setting.
Failure of serial connection. Please check the cable
connection.
Communication format for high level language
program error. Please check if the serial port
number, communication format are correct and
the serial transmission rate is the same as the
device setting.
If the device is resolving the last command, the
host can’t accept the response of the device.
Please try again later.
Steps
1.
Press
SYSTEM
key below the LCD
screen to enter [SYSTEM CONFIG]
setting page.
2.
Use up and down arrow keys to
select
HAND SHAKE
on this
setting page. The selected item will
be highlighted.