GBM-3080/3300 User Manual
166
:SYSTem:CALibration:AUTO
Description
Sets or returns the status of self-calibration
function.
Syntax
:SYSTem:CALibration:AUTO {OFF/0, ON/1}
Query Syntax
:SYSTem:CALibration:AUTO?
Parameter/Return
parameter
0
Disable the self-calibration function of device.
1
Enable the self-calibration function of device.
The device will self-calibrate once per 30
minutes.
Example
->:SYST:CAL:AUTO OFF
->:SYST:CAL:AUTO?
->OFF
:SYSTem:RESult
Description
Sets or returns the means for sending data.
Automatic or by using FETCH instruction.
Syntax
:SYSTem:RESult {FETCH, AUTO}
Query Syntax
:SYSTem:RESult?
Parameter/
Return parameter
FETCH The data can only be returned to the master
through Fetch instruction. The device
passively sends the data.
AUTO
The test results are automatically sent to
the master after each test is completed.
Example
->:SYST:RES AUTO
->:SYST:RES?
->AUTO
:SYSTem:DATAout
Description
Sets or returns the means for sending data.
Set
Query
Set
Query
Set
Query
Summary of Contents for GBM-3080
Page 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Page 21: ...GETTING STARTED 21 Carry position ...
Page 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Page 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Page 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Page 183: ...APPENDIX 183 Dimensions ...