COMMAND OVERVIEW
139
:VOLTage:LiMiT:NOMinal
Description
Sets and returns the nominal value for the voltage,
nominal value is only operated in ABS and PER
mode.
Syntax
:VOLTage:LiMiT:NOMinal<float> -303 ~ 303
Query Syntax
:VOLTage:LiMiT:NOMinal?
Parameter /
Return parameter
Floating
point
Floating point of nominal value. It’s unit
is
V.
Example
->:VOL:LMT:NOM 12.345m
->:VOL:LMT:NOM?
->+12.3450E-3
Note
If the current voltage range is set to NOM and the
comparison mode for voltage is PER or ABS, the
voltage range will be shifted to the optimum range
based on the nominal value.
However, when the comparison mode for voltage is
set to SEQ, the range will not be shifted according to
the nominal value. (range is selected according to the
upper limit in SEQ mode.)
:VOLTage:LiMiT:SEQ
Description
Sets and returns the upper and lower limits for the
voltage on the current readout compassion mode.
Syntax
:VOLTage:LiMiT:SEQ<lower, upper> -303 ~ 303
Query Syntax
:VOLTage:LiMiT:SEQ?
Parameter /
Return parameter
Lower
The lower limit of floating point
upper
The upper limit of floating point
Example
->:VOLT:LMT:SEQ 1.23456, 3.45678
->:VOLT:LMT:SEQ?
->+1.0, +3.0
Set
Query
Set
Query
Summary of Contents for GBM-3080
Page 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Page 21: ...GETTING STARTED 21 Carry position ...
Page 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Page 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Page 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Page 183: ...APPENDIX 183 Dimensions ...