SYSTEM CONFIGURATION
77
Note
All settings on the system configuration page will be
automatically saved in the system and automatically
loaded at the next boot.
Configuring account number
Steps
1.
Press the
Measure
or
Setup
key on
the front panel of the GBM-3300.
or
2.
Press
SYSTEM
key below the LCD
screen to enter [SYSTEM CONFIG]
setting page.
3.
Use up and down arrow keys to
select
ACCOUNT
on this setting
page. The selected item will be
highlighted.
4.
Press option key on the right of the
LCD screen to change setting.
Available
parameter
ACCOUNT
ADMIN, USER
PASSWORD
CHANGE PASSWORD, DELETE
PASSWORD