COMMAND OVERVIEW
153
Note
This command has the same function as
VOLT:LMT:MODE ABS. This command is valid only
for voltage.
:CALCulate:STATistics[:STATe]
Description
Sets or returns the status of processing data.
Syntax
:CALCulate:STATistics[:STATe]{LOG, STAT}
Query Syntax
:CALCulate:STATistics[:STATe]?
Parameter /
Return parameter
LOG
Enable the data logging function
STAT
Enable the statistics function
Example
->:CALC:STAT LOG
->:CALC:STAT?
->LOG
Note
Please use command “LOG:START ON” to enable
data logging and statistic function when trigger mode
is set to INT.
Please use trigger key to enable data logging and
statistic function when trigger mode is set to EXT.
:CALCulate:STATistics:RESistance:NUMBer?
Description:
Queries the numbers of statistics for resistance.
Query Syntax
:CALCulate:STATistics:RESistance:NUMBer?
Return parameter <Total No.>,<Effective No.> Both numbers are
integers.
Example
->:CALC:STAT:RES:NUMB?
->10, 8
Note
Effective quantity doesn’t include the number of
overflow (OF) or error (FAULT). The value is regarded
as valid as long as it can be displayed on the screen.
Set
Query
Set
Query
Summary of Contents for GBM-3080
Page 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Page 21: ...GETTING STARTED 21 Carry position ...
Page 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Page 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Page 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Page 183: ...APPENDIX 183 Dimensions ...