GBM-3080/3300 User Manual
150
Query Syntax
:CALCulate:LIMit:VOLTage:MODE?
HL
Upper and lower limits of direct readout
comparison. (SEQ)
REF
Percentage comparison. (% relative
deviation)(PER)
ABS
Absolute deviation comparison (
Δ
)
Example
->:CALC:LIM:VOLT:MODE HL
->:CALC:LIM:VOLT:MODE?
->HL
Note
:CALCulate:LIMit:VOLTage:UPPer
Description
Sets and returns upper limit for voltage
comparator.
Syntax
:CALCulate:LIMit:VOLTage:UPPer<0 ~ 999999>
Query Syntax
:CALCulate:LIMit:VOLTage:UPPer?
Parameter /
Return parameter
0~999999 Positive integer. If the parameter is more
than 999999, the parameter will be set to
999999 automatically.
Example
->:CALC:LIM:VOLT:UPP 123456
->:CALC:LIM:VOLT:UPP?
->123456
Note
The number of decimal digits and unit are related to
the current range no. Please refer to the following
table for their relationship.
Range no. Number of decimal digits Unit Description
0(8V)
5
V
123456=1.23456V
1(80V)
4
V
123456=12.3456V
2(300V
)
3
V
123456=123.456V
Set
Query