25 V AC
25 V AC
fuse 0,5 AT
25 V AC
Transformer *
Ulveco DK 13963
100-120 V AC 60 Hz 40VA
* or transformer
Guldmann DTS 2420
Note: No changeable fuse
©
Guldmann
GB
-1394/11/08
4
4.
e
le
ct
rI
ca
l
co
n
n
ec
tI
o
n
Fixed rail systems
Fixed rail systems must be connected by Guldmann
Service Team or by a qualified installer .
A Guldmann class I transformer must be used .
Installation to follow existing rules for authorized
electricians and according to UL/EN 60601-1 .
Protective ground must be available and connected
correctly . From the ”connecting point/neutral point”
protective ground to rail maximum permissible resistance
is 0 .2 Ohm .
See ”Instructions: Connecting protective ground to rail” .
See wiring diagram: ”Classification of Guldmann Hoist
System Class I – Permanent installation with protective
ground on rails” .
Classification of Guldmann Hoist System Class I
– Permanent installation with protective ground on rails
Terminals on transformers Class I
Double-insulated UL approved cable must be used
(tested for voltage over 400 Volts) .
Transformer Guldmann DTS-2420
In:
Terminal 1
= 0
Terminal 2
= 100-120 Volt AC (USA)
Terminal 3
= 230 Volt AC
Terminal 4
= protective ground
Out:
Terminal 5
= 25V AC
Terminal 6
= 25V AC
Transformer Ulveco Dk 16
In:
Fuse
= 0,5 AT
Terminal 1
= 0
Terminal 2
= 100-120 Volt AC (USA)
Terminal 3
= protective ground
Out:
Terminal 4
= 25V AC
Terminal 5
= 25V AC
Equipment is disconnected from Supply Mains by breaking
the main breaker switch .
Connection to the supply network
Class I
Summary of Contents for GH2
Page 2: ......
Page 4: ......
Page 10: ...Bla bla Guldmann GB 1394 11 08 12 ...
Page 34: ...Bla bla Guldmann GB 1394 11 08 36 ...
Page 73: ...Bla bla Guldmann GB 1394 11 08 75 4 electrical connection Interconnection diagram GH2 twin ...
Page 87: ...Bla bla Guldmann GB 1394 11 08 89 ...
Page 88: ... Guldmann GB 1394 11 08 90 5 control Control 5 Checkliste 88 ...
Page 90: ... Guldmann GB 1394 11 08 92 6 toolbox ToolS 6 Overview of tools for rail installation 93 ...
Page 93: ... Guldmann GB 1394 11 08 95 7 service GH2 Service manual ...
Page 107: ... Guldmann GB 1394 11 08 109 7 service Periodic Inspection ...
Page 108: ......
Page 109: ......