7.
s
er
vI
ce
Mounting of bracket for the printed circuit board
Only when the strap tightener has been pulled down,
the bracket for the printed circuit board can be mounted
without the micro switch on the back of the printed circuit
board being jammed behind the strap tightener .
Note:
Make sure that the hooks at the end of bracket of the
printed circuit board have engaged with the side plate
before the front hook is engaged .
Inspecting the strap tightener
Check through spyhole in the bracket that the lever of the
micro switch rests
on top
of the strap tightener . Check the
function of the micro switch by tightening and slackening
the strap . When pulling the strap lightly the micro switch
should make a click .