PA5000 Series Power Analyzer Start Guide
When the cut-off frequency (fc) is 1MHz
Voltage/current
0.1Hz ~ fc/10: (f/1000kHz)
2
× 60% of the reading
DC: 0.05% of the range
Power
0.1Hz ~ fc/10: (f/1000kHz)
2
× 120% of the reading
DC: 0.1% of the range
8.3.5
Angle Error
Input waveform: 50Hz sine wave. Common mode voltage: 0V. Line filter: OFF.
Data update rate: 500ms.
±
[|
φ -cos
-1
(
λ
∕
1.0002)|+0.01]deg
Voltage and current are rated range.
8.3.6
Temperature Coefficient
Temperature coefficient: should be added ± 0.0 3%/°C of the reading.
8.3.7
Accuracy of 12 months
Accuracy of 12 months: accuracy of 6 months plus (the reading error for accuracy of 6
months X 0.5).
8.4 Measurement Modes
Table 8-16: Measurement modes
Measurement modes
Description
Normal Mode
The voltage, current, power, waveform computation and integral
values are measured in this mode. You can use the waveform
display X 8, bar graph display X 8, vector graph display X 2 and X-Y
figure X 8.
Harmonic Mode
Harmonics up to 255th order can be measured for a 1kHz
fundamental frequency signal. Please use this function when
conducting harmonic measurement on the signal whose fundamental
frequency is higher than the commercial power frequency.
Harmonics display X 3.
IEC Harmonic Mode
This mode performs harmonics measurement in accordance with
IEC61000-3-2 and IEC61000-4-7 international standards.