6.8 U-I Phase
The delay error is introduced when the external sensor/transformer is connected. The
power analyzer provides the U-I phase setting for phase delay compensation and
phase angle measurement error decrease, thus improving the measurement accuracy.
The U-I phase can be set in the range of -50000.00ns~50000.00ns, and its minimum
resolution is 0.01ns. For the specific setting value, please refer to the specifications of
sensor/transformer.
Figure 6-26: U-I phase setting