6.6.2
Steps
1. Display average processing menu
Press the
“Measure” button to pop up the measurement menu; and then click the “AVG
Setting
” soft key in the menu to display the dialog box shown in Figure 6-24.
Figure 6-24: Average setting menu
2. Configure parameters for averaging function
Finish the following configuration in the menu shown in Figure 6-24.
①
Enable the averaging function.
②
Select the average type.
③
If the average type is the Exp (exponential average), the attenuation constant is set;
if the average type is Lin (moving average), the average number is set.
6.7 Synchronous Measurement
The alone, master or slave mode can be selected in the synchronous measurement.
The alone mode is selected for a single instrument. When two instruments are
connected together, the synchronous master/slave measurement can be performed.
Figure 6-25 shows the synchronous measurement menu.
Figure 6-25: Synchronous measurement menu
Only one instrument should be set as master and one instrument is set as salve. The
external trigger output of master is connected to the external trigger input of salve via
the BNC cable. The master issues the measurement starting signal and the slave
receives the signal, so that the synchronous measurement of two instruments is
achieved.
The slave starts measuring while the master starts measuring.
The slave stops measuring while the master stops measuring.