English (GB)
18
5.7.5 12 V
5.7.6 Analog signals
+12 V
GND
IN
Description
Function
43
-------
44
Batch tank ClO
2
overflow
NC
Contact open in case of overflow
ClO
2
level above contact
45
-------
46
Batch tank ClO
2
max.
NO contact
Contact closed in case of "Max"
ClO
2
level above contact
47
-------
48
Batch tank ClO
2
min.
NO contact
Contact open in case of "Min"
ClO
2
level below contact
49
-------
50
Batch tank ClO
2
dry run
NO contact
Contact open in case of dry run
ClO
2
level below contact
51 (brown)
-------
52 (white)
Pre-empty signal NaClO
2
NC
Contact open in case of pre-empty signal
NaClO
2
level below contact
-------
-------
53 (green)
Empty signal NaClO
2
NC
Contact open in case of empty signal
NaClO
2
level below contact
-------
-------
54 (green)
Empty signal HCl
NC
Contact open in case of empty signal
HCl level below contact
55 (brown)
-------
56 (white)
Pre-empty signal HCl
NC
Contact open in case of pre-empty signal
HCl level below contact
57
-------
58
-----------------------------------
-----------------------------------------
71 (white)
72 (brown)
73 (green)
Hall sensor NaClO
2
pump
75 (white)
76 (brown)
74 (green)
Hall sensor HCl pump
77 (white)
78 (brown)
79 (green)
-----------------------------------
-
81
80
83
-
84
Contact input water meter
85
-
86
Remote on/off
Contact open: "Remote off"
87
-
88
Fault gas warning device
Contact open: fault
89
-
90
Contact main water min.
Contact open: min. contact main water
Shield
IN/OUT
Description
95
93 IN +
Analog IN
96
97 OUT +
Analog OUT
Summary of Contents for Oxiperm OCD-164
Page 2: ...2...
Page 247: ...Fran ais FR 247 4 crou filtre PVC gris...
Page 422: ...RU 422 10 10...
Page 431: ...RU 431 10 5 5 2 1 5 5 6 5 6 1 5 6 1 TM04 8202 4510 50 60 6 20 11 1D O2 O...
Page 432: ...RU 432 5 6 2 1 2 3 20 20 4 11 4 5 4 4 5 4 6 5 6 5 6 TM04 8203 4510 1 20 2 3 4 5 6...
Page 526: ...526...