English (GB)
40
9.3 Maintenance of reactor
9.3.1 Spare valves (3 per reactor)
Fig. 27
Reactor with valves
Warning
Prior to any maintenance work on the reactor, rinse
the system with water until no more chemicals are
present.
Check the valves on the reactor for leaks each time
maintenance work is carried out. Replace valves if
necessary.
Note
For reasons of safety, no spare parts are offered for
reactors, apart from the valves.
In case of faults, the entire reactor must be
exchanged.
OCD-164
System pressure
Product No.
-30 to -220
< 3 bar
91836549
(553-640)
> 3 bar
96689071
(553-641)
-350 to -2000
< 3 bar
91836550
(553-642)
> 3 bar
96689073
(553-643)
Note
The reactor valves are not identical to the valves of
the dosing pumps.
Do not interchange the valves!